Measurement system analysis glossary
The main figures and terms of MSA, VDA 5 and process capability, explained briefly, with formula and a link to the detailed article. To calculate: Cg/Cgk calculator, gage R&R calculator, Cpk calculator and free Excel templates.
AV, reproducibility
Appraiser variation: how much the results of different appraisers on the same part differ.
Bias
Measured mean minus reference value. It shows whether a gage reads too high or too low on average.
More: MSA 1: Type 1 gage study, calculating Cg and Cgk with example
Cg
Capability index of the type 1 gage study: it compares the spread of a gage on a reference standard with 20 % of the tolerance. At least 1.33 is common.
More: MSA 1: Type 1 gage study, calculating Cg and Cgk with example
Cgk
Like Cg, but also takes the bias into account, i.e. how far the mean lies from the reference value. Cgk is never larger than Cg.
More: MSA 1: Type 1 gage study, calculating Cg and Cgk with example
Cm and Cmk
Machine capability indices: spread of a machine in a short run relative to the tolerance, Cmk also with the location.
Control chart
Chart over time with a centre line and control limits. Points beyond the limits or unusual patterns show that something has changed.
More: MSA 4: Measurement system stability study with control chart and example
Coverage factor k
Factor that turns the combined standard uncertainty into the expanded uncertainty U. k = 2 is common, which corresponds to about 95 %.
More: Measurement uncertainty explained: type A, type B and k = 2
Cp and Cpk
Process capability indices using the variation within subgroups. Meaningful only if the process is stable; otherwise Ppk applies.
More: Cpk calculator
EV, repeatability
Equipment variation: the spread when the same appraiser measures the same part several times. It comes mainly from the gage and the fixture.
Gage capability
Whether a gage is suitable for a given tolerance. It is usually shown with a type 1 study (Cg, Cgk) and a gage R&R (%GRR, ndc); also written gauge capability.
More: MSA 1: Type 1 gage study, calculating Cg and Cgk with example
%GRR
Share of the measurement system variation GRR in the tolerance or in the total variation. Common: up to 10 % acceptable, up to 30 % conditionally acceptable.
Kappa κ
Measure of agreement of good/bad decisions, corrected for the share that would agree by chance. According to AIAG, above 0.75 is good.
More: MSA 6: Attribute agreement analysis with kappa, with example
Linearity
Shows whether the bias changes across the measuring range. Several reference standards are measured and a line is fitted through the deviations.
Measurement uncertainty
The range around a measured value in which the true value lies with high probability. The individual contributions are combined as a root sum of squares.
More: Measurement uncertainty explained: type A, type B and k = 2
ndc
Number of distinct categories: how many classes the measurement system reliably separates within the part variation. Usually at least 5 are required.
Pp and Ppk
Process performance indices: spread of all individual values of a process relative to the tolerance, Ppk also with the location.
More: Cpk calculator
PV, part variation
How much the parts of a gage R&R really differ. For a meaningful result the parts should cover the variation of production.
QMP
Capability ratio of the measuring process under VDA 5: additionally with appraisers, parts and environment. At most 30 % is recommended.
More: VDA 5: Measurement uncertainty of measurement system and process, with example
QMS
Capability ratio of the measurement system under VDA 5: the expanded uncertainty on the reference standard relative to the tolerance. At most 15 % is recommended.
More: VDA 5: Measurement uncertainty of measurement system and process, with example
Resolution, %RE
The smallest step a gage displays, relative to the tolerance. At most 5 % is common under VDA 5; the AIAG manual names the rule of tens.
Signal detection
Attribute study with measured reference values: it determines the width d of the zones around the limits in which decisions vary, and from it a %GRR.
More: MSA 7: Attribute gage study with signal detection and acceptance curve
Stability
Shows whether a measurement system changes over time, for example through drift or wear. The same part is measured regularly and tracked on a control chart.
More: MSA 4: Measurement system stability study with control chart and example
Test equipment capability
Another word for gage capability, often used for the whole inspection process with appraisers and parts. Bosch Booklet 10 speaks of the capability of measurement and test processes.
More: Bosch Booklet 10: Capability of measurement and test processes explained
Tolerance T
The permitted range of a characteristic, upper minus lower limit. Capability indices always apply to the tolerance used in the calculation.
More: MSA 1: Type 1 gage study, calculating Cg and Cgk with example
TV, total variation
Measurement system variation and part variation together. %GRR refers to it when no tolerance is used as reference.