Measurement uncertainty explained: type A, type B and k = 2
What measurement uncertainty means, how to estimate it from a few inputs and why the largest contribution determines almost everything. Without heavy maths, with a small budget to try out.
- A result is measured value ± U, with k stated.
- Type A from measurement series, type B from certificates, limits, resolution.
- Contributions combine as root sum of squares; the largest dominates.
- k = 2 covers about 95 %.
Contents
An everyday example
Imagine weighing yourself on the bathroom scale in the morning: 78.4 kg. You step off and on again: 78.6 kg. A third time: 78.3 kg. Which value is right? None exactly. The scale displays one decimal place, but its values vary by a few hundred grams. Also, you do not know whether the scale is set correctly. Perhaps it always shows half a kilo too little.
So the honest statement would be: “I weigh about 78.4 kg, plus or minus half a kilo.” This “plus or minus” is the measurement uncertainty. In production it is the same, only it is about micrometres instead of kilograms, and the question is whether a part lies within the tolerance.
What is measurement uncertainty?
No measured value is exact. The measurement uncertainty states the range in which the true value lies with high probability. A complete result therefore consists of the measured value and the uncertainty, for example “10.0412 mm ± 2.9 µm (k = 2)”. How to get there is described in the GUM, the international Guide to the Expression of Uncertainty in Measurement.
Type A and type B
Every influence is expressed as a standard uncertainty u, i.e. like a standard deviation. The GUM distinguishes two ways of obtaining it, which are equally valid:
- Type A: statistically from repeated measurements, such as the standard deviation s of 25 measurements on a standard.
- Type B: from other knowledge, such as the calibration certificate, manufacturer specifications, resolution or experience.
Type B needs an assumption about the distribution:
| Given as | Standard uncertainty |
|---|---|
| Calibration certificate: U with coverage factor k | u = U / k |
| Limit ± a, every value equally likely (rectangle) | u = a / √3 |
| Resolution RE of a display | u = RE / (2 √3) |
Where do these formulas come from? Three short explanations:
- Calibration certificate: the certificate usually states the expanded uncertainty U with k = 2, i.e. already “times 2”. The budget needs the simple standard uncertainty, so divide by k. With U = 1.0 µm and k = 2, u = 0.5 µm.
- Limit ± a: if you only know “at most ± 1.5 µm” but not where exactly in between, you assume every value is equally likely. The standard deviation of such a uniform distribution is a divided by √3 ≈ 1.73, i.e. about 58 % of a. With a = 1.5 µm, u = 1.5 / 1.73 = 0.87 µm.
- Resolution: a display with steps of 1 µm rounds. The true value lies up to half a step away, i.e. ± 0.5 µm, equally likely anywhere. That is a limit with a = RE / 2, hence u = RE / (2 √3) = 0.29 µm.
How to calculate measurement uncertainty: step by step with formula and example
Independent influences do not simply add up; they combine as the root of the sum of squares. As a result, the largest contribution almost alone determines the result, and small contributions hardly matter.
An example: a diameter is measured with a micrometer, result 10.0412 mm. Four influences are known: the setting master (calibration certificate U = 1.0 µm, k = 2), the repeatability (standard deviation s = 0.97 µm from 25 measurements), the temperature (estimated at most ± 1.5 µm) and the resolution (1 µm). Calculation:
| Influence | Calculation | u in µm | u² in µm² |
|---|---|---|---|
| Calibration | 1.0 / 2 | 0.500 | 0.250 |
| Repeatability | s | 0.970 | 0.941 |
| Temperature | 1.5 / √3 | 0.866 | 0.750 |
| Resolution | 1 / (2 √3) | 0.289 | 0.083 |
| Sum | 2.024 |
So the complete result is 10.0412 mm ± 2.9 µm (k = 2). How much does each influence contribute? Divide its u² by the sum:
| Influence | Type | u in µm | Share of uc² |
|---|---|---|---|
| Calibration (certificate: U = 1.0 µm, k = 2) | B | 0.500 | 12 % |
| Repeatability (s from 25 measurements) | A | 0.970 | 46 % |
| Temperature (limit ± 1.5 µm) | B | 0.866 | 37 % |
| Resolution 1 µm | B | 0.289 | 4 % |
uc = 1.42 µm, U = 2 · uc = 2.85 µm. Repeatability and temperature together account for over 80 %; a finer resolution would change practically nothing.
VDA 5 calculates slightly differently: of resolution and repeatability only the larger counts, because the spread of the values already contains the resolution. More in VDA 5.
What U is used for
- State the result: measured value ± U with k; U with at most two significant digits, rounded up.
- Decide conformance: with a guard band of U at each limit, as in VDA 5, conformance is proven if the measured value is at least U away from the limit. The agreed decision rule applies (ISO 14253-1).
- Assess measurement methods: VDA 5 relates U to the tolerance (QMS, QMP).
- Find the biggest lever: the budget shows where an improvement helps and where it does not.
Download the free versionFree MSA software: what STATlum Free does
Common mistakes
- Putting the expanded uncertainty from the calibration certificate directly into the budget. Divide by k first, otherwise it counts twice.
- Not dividing limits by √3. A limit ± a is not a standard uncertainty. Without the conversion the contribution is 73 % too large.
- Simply adding everything. Independent contributions are added in quadrature.
- Forgetting influences. What is not in the budget makes the uncertainty artificially small. Temperature and part form are overlooked particularly often.
- Not correcting a known systematic deviation. If a gage is known to read 2 µm too high, correct it; only the uncertainty of this correction belongs in the budget.
- Stating U without k. “± 2.9 µm” alone does not say how certain the statement is. Always state k.
Frequently asked questions
Is measurement uncertainty the same as measurement error?
No. The error is the difference from the true value and usually unknown; a known systematic error is corrected. The uncertainty describes how well the result is known.
Why k = 2?
For a normal distribution ± 2 u covers about 95 % of the values. k = 2 is the usual convention, also on calibration certificates. With few measurements a larger k from the t distribution may be needed.
Is type B less accurate than type A?
Not in principle. A well-founded type B value can be more reliable than a type A value from few measurements.
Sources
- JCGM 100:2008 (GUM): Evaluation of measurement data, Guide to the expression of uncertainty in measurement
- NIST Technical Note 1297: Guidelines for Evaluating and Expressing the Uncertainty of NIST Measurement Results
- DKD-L 13-1: Guideline on determining measurement uncertainty (PTB, German)
- ISO 14253-1: Geometrical product specifications, inspection by measurement, decision rules
- Fluke: Getting started with uncertainty analysis (German)
This article explains the method. It does not replace the guideline that applies to your part or the professional assessment of an actual study.