VDA 5: Measurement uncertainty of measurement system and process, with example
- VDA 5 assesses a measurement method by its measurement uncertainty: every influence gets a standard uncertainty.
- First the measurement system on the standard (Q_MS at most 15 %), then the measurement process with appraisers, parts and environment (Q_MP at most 30 %).
- To prove conformance, the usable tolerance shrinks by a guard band at each limit.
Contents
What this is about
Every measured value is a little bit wrong. That is not a criticism but physics: the standard used to set the gage is itself only known to a certain accuracy, the display rounds, the temperature varies, the part is not perfectly round, and every appraiser applies the gage slightly differently. Honestly, you should therefore not write “10.046 mm” but “10.046 mm ± 0.005 mm”. This ± 0.005 mm is the measurement uncertainty.
VDA 5 asks: how large is this uncertainty, and is it small enough for the tolerance? Unlike classic MSA, which reads spreads from a study, VDA 5 collects all influences individually, like a list of ingredients, rates each with a number and adds them up. That gives an uncertainty budget. You then see not only whether the measuring method is adequate but also which influence costs the most.
The terms in plain words
- Standard uncertainty u
- The uncertainty of a single influence, expressed like a standard deviation. Each influence gets its own u, for example uT for temperature.
- Evaluation from measurements
- Some u are determined from series of measurements, for example repeatability as the standard deviation of 25 measurements.
- Evaluation from prior knowledge
- Other u are estimated from data sheets, calibration certificates or experience, for example “temperature distorts the result by at most 1.5 µm”.
- Rectangular distribution
- If you only know a limit ± a and do not know where the value lies in between, you assume every value is equally likely. The standard deviation of such a distribution is a / √3, i.e. about 0.58 · a.
- Combined uncertainty
- All u are squared, added, and the square root of the sum is taken. That is how independent influences add up, similar to variances in a gage R&R.
- Expanded uncertainty U
- The combined uncertainty times 2. The range ± U contains the true value with about 95 % probability.
- Measurement system MS and process MP
- The measurement system is the instrument with standard and display, checked under the best conditions. The measurement process is measuring in daily use with real parts, appraisers and environment.
- Capability ratio Q
- What percentage of the tolerance the uncertainty takes up: Q = 2 · U / T. Smaller is better.
Purpose
VDA 5 “Measurement and inspection processes” assesses a measurement method by its measurement uncertainty: how far can a result deviate from the true value, and is that small enough for the tolerance? VDA 5 separates two stages: the measurement system and the measurement process.
The separation has a practical reason. If the instrument alone is already too inaccurate, there is no need to investigate further. If the instrument is good but the measurement process is not, the problem lies with parts, appraisers or environment, and that is where to act. The difference from classic MSA is explained in VDA 5 or MSA, the basic idea of measurement uncertainty in Measurement uncertainty explained.
How it works
- Collect influences: what can distort the result? Typical are standard, resolution, repeatability, bias, appraiser, temperature, form of the part. A cause-and-effect diagram helps not to forget anything.
- Measurement system on the standard: determine the calibration uncertainty uCAL from the calibration certificate, resolution uRE, repeatability uEVR from at least 25 measurements on the standard and bias uBI, and the linearity where needed.
- Assess QMS: Dietrich and Radeck recommend at most 15 %. If the measurement system already fails, the second stage is not worth doing.
- Measurement process: also repeatability on the part uEVO and appraiser influence uAV, usually from a gage R&R, plus part, temperature, stability and other influences, often estimated as a limit.
- Assess QMP: recommended at most 30 %.
- Adjust the tolerance: to prove conformance, the tolerance shrinks by a guard band at both limits. The agreed decision rule applies (ISO 14253-1).
Worked example, step by step
A diameter of 10.000 ± 0.050 mm (T = 100 µm) is measured with a micrometer, resolution 0.001 mm, as in MSA 2. Known are:
- According to its calibration certificate, the setting master has xm = 10.0003 mm with an expanded uncertainty U = 1.0 µm (k = 2).
- 25 measurements on the master give the mean x̄ = 10.00076 mm and the standard deviation s = 0.97 µm.
- Repeatability on the part and appraiser influence come from the gage R&R of MSA 2.
- Temperature is estimated to distort the result by at most ± 1.5 µm, the form deviation of the part by at most ± 2.0 µm.
All data are artificial; the measurements on the master are available below as CSV.
Step 1: convert each influence into a standard uncertainty
The influences come in very different forms: as an expanded uncertainty, as a resolution, as a standard deviation, as a limit. To be able to combine them, they are all brought to the same form, the standard uncertainty u.
From the gage R&R come uEVO = 1.190 µm (repeatability on the real part) and uAV = 1.407 µm (appraiser influence); there was no interaction there, uIA = 0. All contributions at a glance:
| Contribution | Influence | u in µm | counted in the process |
|---|---|---|---|
| uCAL | Calibration of standard | 0.500 | yes |
| uRE | Resolution | 0.289 | no (smaller) |
| uEVR | Repeatability on standard | 0.970 | no (smaller) |
| uBI | Bias | 0.266 | yes |
| uEVO | Repeatability on part | 1.190 | yes |
| uAV | Appraiser influence | 1.407 | yes |
| uIA | Interaction | 0.000 | yes |
| uOBJ | Form deviation of part | 1.155 | yes |
| uT | Temperature | 0.866 | yes |
Step 2: measurement system
For the measurement system, calibration, resolution, repeatability on the standard and bias count. But resolution and repeatability overlap: the scatter of the 25 measurements already contains the rounding of the display. So only the larger of the two counts, here uEVR = 0.970 µm. The contributions are squared and added:
| Contribution | Influence | u in µm | u² in µm² |
|---|---|---|---|
| uCAL | Calibration of standard | 0.500 | 0.250 |
| uEVR | Repeatability on standard | 0.970 | 0.940 |
| uBI | Bias | 0.266 | 0.071 |
| Sum | 1.261 |
The capability ratio relates the whole uncertainty range, i.e. 2 · U from −U to +U, to the tolerance:
Step 3: measurement process
For the measurement process, the influences of daily use are added. Of the three repeatabilities (on the standard, on the part, resolution) again only the largest counts, here uEVO = 1.190 µm, because it contains the others. The last column shows how much each contribution adds to the sum of squares:
| Contribution | Influence | u in µm | u² in µm² | Share |
|---|---|---|---|---|
| uCAL | Calibration of standard | 0.500 | 0.250 | 4 % |
| uEVO | Repeatability on part | 1.190 | 1.415 | 24 % |
| uBI | Bias | 0.266 | 0.071 | 1 % |
| uAV | Appraiser influence | 1.407 | 1.980 | 34 % |
| uIA | Interaction | 0.000 | 0.000 | 0 % |
| uOBJ | Form deviation of part | 1.155 | 1.333 | 23 % |
| uT | Temperature | 0.866 | 0.750 | 13 % |
| Sum | 5.799 | 100 % |
Step 4: tolerance adjustment
What does UMP = 4.8 µm mean for the decision on an individual part? A measured value right at the limit could in reality be up to 4.8 µm off, i.e. inside or outside the tolerance. If you want to prove reliably that a part is OK, you therefore subtract UMP at both limits:
A part at 10.046 mm lies within the tolerance, but in the uncertainty range at the upper limit: conformance is not proven. Whether such a part is shipped anyway is governed by the decision rule agreed with the customer.
What the result means
QMS = 4.5 % and QMP = 9.6 % are clearly below the recommended 15 % and 30 %. Measurement system and measurement process are suitable for this tolerance. In words: the uncertainty of a measured value is about ± 5 µm with a tolerance of 100 µm.
The largest contribution comes from the influence “Appraiser influence” with 34 % of the sum of squares. To reduce the uncertainty, start there, for example with a more precise measuring instruction and training, as described in MSA 2. The form deviation of the part is also worth noting: it is not an error of the gage, but it belongs to the measurement process because it influences the result depending on where you measure.
Common mistakes
- Forgetting influences. What is not in the budget does not count, and the uncertainty looks smaller than it is. A cause-and-effect diagram at the start helps.
- Using the expanded instead of the standard uncertainty. If the calibration certificate states U with k = 2, divide by 2 before entering it in the budget.
- Counting twice. Resolution and repeatability, or repeatability on the standard and on the part, overlap. Take only the largest.
- Simply adding standard uncertainties. Correct is: square, add, take the square root.
- Estimating limits too generously. “Temperature ± 10 µm, to be on the safe side” makes the budget useless. Better estimate with reasons and record the reasons.
- Confusing Q with %GRR. Both relate to the tolerance but are calculated differently. The numbers are not directly comparable, see VDA 5 or MSA.
In STATlum
Create a “VDA 5” study on the characteristic. The budget is a table: measurement system blue, measurement process green. Contributions come from measurement series, from limits or as a direct value. The evaluation shows all indices of this example.


Download the free versionExample data as CSVFree MSA software: what STATlum Free doesCalculate a gage R&R online
Frequently asked questions
Which limits apply to Q_MS and Q_MP?
Dietrich and Radeck recommend Q_MS at most 15 % and Q_MP at most 30 %; publications of the VDA on the 3rd edition cite the same values. Your customer’s requirement is binding.
Does VDA 5 replace the gage R&R?
No. In VDA 5 the gage R&R provides the contributions repeatability on the part and appraiser influence. VDA 5 adds further influences such as calibration, temperature and form deviation.
What is the smallest testable tolerance?
The tolerance at which Q just reaches its limit: T_min = 2 · U / Q_max. If a characteristic’s tolerance is smaller, the measurement method is not adequate for it.
What sample size is needed?
For the measurement system at least 25 measurements on the reference standard for repeatability. Repeatability on the part and appraiser influence usually come from a gage R&R. Further influences such as temperature are often estimated as limit values.
Sources
- VDA: Volume 5, Measurement and inspection processes. Capability, planning and management, 3rd edition 2021
- Dietrich, Radeck: Prüfprozesseignung nach VDA 5 und ISO 22514-7 (German), Hanser 2014
- ISO 22514-7: Capability of measurement processes
- Dietrich: Wo liegen die Unterschiede? Vergleich von MSA und VDA Band 5 (German), QZ 6/2011
- Muelaner: VDA-5, Combining Uncertainty Evaluation with Gage Studies
- DKD-L 13-1: Guideline on determining measurement uncertainty (PTB, German)
This article explains the method. It does not replace the guideline that applies to your part or the professional assessment of an actual study.