Cg/Cgk calculator

Paste the measurements, enter the reference value and the tolerance: the calculator determines mean, standard deviation, bias, Cg and Cgk as in a type 1 gage study. The calculation runs only in your browser, nothing is sent or stored.

or limits

Settings

Results appear as soon as measurements, reference value and tolerance are entered.

Commonly required: Cg and Cgk of at least 1.33. What counts is the requirement for your part.

How the calculator works

With the defaults K = 20 % and L = 6 the formulas of the MSA 1 article and of STATlum apply. The standard deviation is calculated with n − 1.

x̄=∑xin
s=∑(xi−x̄)2n−1
B=x̄−xref
Cg=0.2·T6·s
Cgk=0.1·T−|B|3·s

Explained step by step, with a worked example and bias test: Type 1 gage study: calculating Cg and Cgk. With the example data the calculator gives the same values as the article.

Prefer an Excel file: free MSA Excel templates, without macros and with open formulas.

The complete study

Run chart, histogram, bias test, confidence intervals and report are produced by STATlum Free, free of charge and offline on Windows. Download STATlum Free