MSA 1: Type 1 gage study, calculating Cg and Cgk with example
- A type 1 study checks a gage on a calibrated reference standard: spread and bias.
- One appraiser measures the standard 50 times. Cg compares the spread with 20 % of the tolerance, Cgk also subtracts the bias.
- At least 1.33 is common for both. The result applies only to the tolerance used in the calculation.
Contents
What this is about
Imagine a new micrometer arrives at a workstation in your production, and the customer wants proof that it is suitable for measuring a shaft diameter. The micrometer displays values with three decimal places. That looks precise, but it says nothing yet about whether those values can be trusted.
A gage can be off in two ways. First, it can scatter: if you measure the same part several times, you get slightly different values each time. Second, it can be shifted: the values lie close together but are all a little too high or too low. Think of a target. If a shooter scatters, the hits spread over the target. If the sight is misaligned, the hits are close together but off centre.
The type 1 gage study (also written gauge study), also called MSA 1, measures exactly these two things. Instead of a production part it uses a reference standard, a test piece whose true size is known from calibration. Because the true size is known, every deviation can only come from the gage and the measuring itself. At the end there are two indices: Cg for the scatter and Cgk for scatter and shift together.
The terms in plain words
- Gage capability
- Whether a gage is suitable for a given tolerance. In British English the word is often spelled gauge; the type 1 study shows it with Cg and Cgk.
- Reference standard (master)
- A test piece of known size, for example a setting ring or a gauge block. It is calibrated regularly. Its size, the reference value, is stated on the calibration certificate.
- Reference value
- The “true” size of the standard according to the calibration certificate, 25.004 mm in the example. Not the nominal size from the drawing: a standard with a nominal size of 25.000 mm may actually measure 25.004 mm.
- Tolerance T
- The permitted range of the characteristic, i.e. upper limit minus lower limit. For 25.000 ± 0.030 mm that is 25.030 − 24.970 = 0.060 mm.
- Resolution
- The smallest step the gage displays, 0.001 mm for the micrometer. If the step is too coarse, you cannot see the scatter at all.
- Mean x̄
- The sum of all measured values divided by their number. It shows where the gage measures “on average”.
- Standard deviation s
- A measure of how far the individual values typically lie from the mean. A small s means the values are close together.
- Spread 6 s
- Six times the standard deviation, three s below and three s above. For normally distributed values this range contains 99.73 % of all measurements, practically almost all of them.
- Bias
- The shift: mean minus reference value. A bias of +0.7 µm means the gage reads 0.7 µm too high on average.
- Cg and Cgk
- The two capability indices. The larger, the better. A minimum of 1.33 is common.
Purpose
The type 1 gage study checks whether a gage measures a calibrated reference standard repeatably and without bias: how much the values scatter (repeatability) and how far their mean lies from the reference value (bias).
It is usually the first step of a measurement system analysis. The idea behind it is simple: if the gage alone, under the best conditions and with a standard, is already too imprecise, it will not get better with real parts and different appraisers. Then no further study is worthwhile, and the gage has to be dealt with first.
What the type 1 study does not check: the influence of different appraisers and the variation of real production parts, which is what MSA 2 is for. Whether the gage measures equally well over its whole range is shown by linearity in MSA 5, whether it stays the same over weeks by stability in MSA 4.
How it works
- Check the resolution. The display must be fine enough, at most 5 % of the tolerance. For T = 0.060 mm that is 0.003 mm, so a micrometer with 0.001 mm is fine enough. More in Gage resolution.
- Choose the standard. It should be calibrated and as close as possible to the middle of the tolerance, because that is where most parts will lie later. Take the reference value from the calibration certificate, not from the drawing.
- Measure 50 times, at least 25 times. Measure as in production: same place, same fixture, same appraiser. Take the standard out and reload it every time, because loading is part of measuring. Write down all values unrounded and do not leave any out, not even an “outlier”.
- Calculate. From the 50 values determine mean, standard deviation and bias, then Cg and Cgk. The example below shows every step.
- Assess. It is common to require Cg and Cgk of at least 1.33; some customers require 1.67. What counts is always the requirement for your part.
Worked example, step by step
An appraiser measures a setting master for a shaft diameter of 25.000 ± 0.030 mm 50 times with a micrometer. According to the calibration certificate the standard is 25.004 mm. The data are artificial and available below as CSV, so you can recalculate every step yourself, for example in Excel.
Step 0: input data
| Item | Value |
|---|---|
| Lower limit LSL | 24.970 mm |
| Upper limit USL | 25.030 mm |
| Tolerance T = USL − LSL | 0.060 mm = 60 µm |
| Reference value of the standard | 25.004 mm |
| Resolution | 0.001 mm = 1.7 % of tolerance |
| Number of measurements n | 50 |
From here on we calculate in micrometres (µm, thousandths of a millimetre). That makes the numbers easier: 0.060 mm is 60 µm.
Before calculating, it is worth looking at this run chart. It should look like a calm, random band. If the values rise steadily, something may be warming up; if they suddenly jump to another level, something has shifted. Then find the cause first, because the calculation assumes that nothing changed during the measurement.
Step 1: mean
Add up all 50 values and divide by 50. The sum is 1250.234 mm.
Step 2: bias
Step 3: standard deviation
The standard deviation says how far the values scatter around the mean. It is calculated like this: for each value find its deviation from the mean, square that deviation, add up all squares, divide by n − 1 = 49 and take the square root. Because only six different values occur in this example, they can be summarised in a table:
| Value in mm | Count | Deviation in µm | Square in µm² | Count · square |
|---|---|---|---|---|
| 25.002 | 2 | -2.68 | 7.182 | 14.365 |
| 25.003 | 8 | -1.68 | 2.822 | 22.579 |
| 25.004 | 11 | -0.68 | 0.462 | 5.086 |
| 25.005 | 16 | 0.32 | 0.102 | 1.638 |
| 25.006 | 9 | 1.32 | 1.742 | 15.682 |
| 25.007 | 4 | 2.32 | 5.382 | 21.530 |
| Total | 50 | 80.880 |
Why square? Otherwise deviations upwards and downwards would cancel each other out, and large deviations should count more than small ones. Why n − 1 rather than n? The mean was calculated from the same data, so the values lie on average a little closer to it than to the true mean. Dividing by n − 1 compensates for that.
Step 4: Cg
For Cg a band of 20 % of the tolerance, i.e. ± 10 %, is placed around the reference value. With a tolerance of 60 µm the band is 12 µm wide. The question is: how many times does the spread of the gage fit into this band? If it fits exactly once, Cg = 1. To leave a reserve, 1.33 is usually required, so the spread may take up at most three quarters of the band.
Step 5: Cgk
Cg only looks at the width of the scatter, not at where it lies. Cgk also takes the bias into account. Because the mean is shifted upwards by 0.68 µm, it gets tighter on the upper side. So the calculation uses half the band (6 µm) minus the bias and compares that with half the spread (3 · s).
Recalculate with your own measurements: the Cg/Cgk calculator calculates the same way, right in the browser.
Step 6: is the bias real or chance?
Even a perfect gage would not have exactly the reference value as its mean after 50 measurements; there is always a bit of chance. The bias test checks whether the shift of 0.68 µm is larger than chance alone would explain. To do this, the bias is divided by the uncertainty of the mean, s / √n. This uncertainty is much smaller than s itself, because chance largely evens out when many values are averaged.
Roughly speaking: if t is larger than about 2, the shift is no longer chance. The p-value says the same more precisely: it is the probability of seeing such a large shift although the gage is actually not shifted at all. Below 0.05 the bias is usually regarded as significant.
Step 7: how certain is the result?
Cg and Cgk are estimated from 50 values. If the study were repeated, the result would be slightly different. The confidence interval states the range in which the true value lies with 95 % certainty.
What the result means
Cg = 1.56 and Cgk = 1.38 are both above 1.33. In words: the micrometer scatters little enough and is only slightly shifted. It is capable for this characteristic with this tolerance.
Two things should still be noted. First, the bias is real, even if it is small; that is a reason to look at the setting and calibration of the gage. Second, the confidence interval of Cgk reaches down to 1.09, i.e. below 1.33. The margin to the limit is small. Even a slight deterioration, for example through wear, can tip the result.
And: the result applies only to this tolerance. The tolerance is part of the formula; the gage stays the same. With a tighter tolerance of 0.030 mm the same micrometer would reach only Cg = 0.78 and Cgk = 0.60. The smallest tolerance for which the gage still suffices is 0.058 mm.
| Result | Meaning | What to do |
|---|---|---|
| Cg and Cgk ≥ 1.33 | gage scatters little and hits the centre | continue with MSA 2 or MSA 3 |
| Cg ≥ 1.33, Cgk too small | scatter fine, but shifted | reset or adjust the gage, check the reference value |
| Cg and Cgk too small | gage scatters too much | look for causes: fixture, probing, temperature, the gage itself |
If an index is too small, see Measurement system not capable.
Common mistakes
- Reference value from the drawing instead of the calibration certificate. Then the bias is measured against a wrong value, and Cgk comes out too good or too bad.
- Standard not reloaded. If the standard stays in place and is only probed 50 times, the scatter from loading is missing. Cg then looks better than it is in production.
- Values rounded or “outliers” removed. Both make the scatter artificially smaller. A conspicuous value points to a cause; it is not an error in the list.
- Too few measurements. With 10 or 15 values the result is a matter of luck, as the second chart above shows.
- Transferring the result to another tolerance. A gage that is capable for ± 0.030 mm may no longer be capable for ± 0.015 mm.
- Ignoring temperature. A standard that has just come from a cold store grows during the measurement. Let standard and gage adjust to the room first.
In STATlum
Create an “MSA 1” study on the characteristic, enter tolerance, reference value and resolution, and type in or import the measurements. The evaluation shows all indices of this example, plus histogram and smallest tolerance.


Download the free versionExample data as CSVVideo on MSA 1Free MSA software: what STATlum Free doesCalculate Cg and Cgk onlineFree MSA Excel templates
Frequently asked questions
How many measurements does a type 1 study need?
Usually 50 measurements on one standard, at least 25. With fewer measurements the confidence interval of Cg and Cgk becomes much wider.
What is the difference between Cg and Cgk?
Cg assesses only the spread of the values. Cgk also takes into account how far the mean deviates from the reference value. Without bias both are equal.
Which limit applies to Cg and Cgk?
Often 1.33; some customers require 1.67. The requirement of your customer or company is binding.
Is the type 1 study part of the AIAG MSA manual?
No. Cg and Cgk come from company guidelines; the AIAG manual treats bias and repeatability with other indices.
What is the difference between a type 1 study and gage R&R?
A type 1 study checks the gage alone on a reference standard: one appraiser, repeated measurements, spread and bias. A gage R&R measures real production parts with several appraisers and also shows the appraiser influence. The type 1 study usually comes first.
Sources
- Robert Bosch GmbH: Booklet 10, Capability of Measurement and Test Processes, edition 11.2019
- AIAG: Measurement Systems Analysis, Reference Manual, 4th edition
- Minitab: Methods and formulas for the type 1 gage study
- NIST/SEMATECH e-Handbook: Analysis of bias
This article explains the method. It does not replace the guideline that applies to your part or the professional assessment of an actual study.