Measurement system not capable: find the causes and improve
A “not capable” result is the start of the work, not the end. How to check whether the study is right, where the variation comes from and which levers work.
- First check the study: tolerance, reference value, resolution, part selection, data.
- The weak index points the way: EV, AV, bias, stability, kappa, budget.
- The mean of n measurements only lowers repeatability, and only if production does the same.
- Do not massage the result; if in doubt, approve conditionally with documentation or change the method.
Contents
First of all: no reason to panic
The study has been evaluated, and there it is: “not capable”. That feels like a bad grade. But it is not. The result only says that the measurement system in its current form brings too much variation for this tolerance. It does not say that the gage is scrap, nor that anyone did anything wrong.
Such a result is more of a diagnosis. And as with a doctor, a diagnosis only helps if you look more closely: where exactly does it hurt? The indices of a measurement system analysis provide a surprising number of clues for this. This article shows how to read them and in which order to proceed.
- Check the study: was the result obtained correctly at all?
- Narrow down the cause: which index falls short, and what does that say about the cause?
- Apply the lever: start with the largest contribution, not the most convenient one.
- Repeat: after each measure, repeat the study and compare.
Check the study first
A poor result is often not caused by the gage but by the study. Before investing, check these points:
- Was the correct tolerance used, and does the reference value match the calibration certificate?
- Is the resolution fine enough (below 5 % of the tolerance)?
- Is the gage calibrated and set correctly?
- Do the parts cover the production variation, or do they all come from the middle?
- Are the data consistent: single outliers, swapped parts, an appraiser who works differently?
More on resolution in Gage resolution.
Where does the variation come from?
The index that falls short usually already points the way:
| Finding | Likely causes | Actions |
|---|---|---|
| Cg too small | Fixture, probing, measuring force, temperature, dirt | improve the fixture, define the force, stabilise temperature, clean |
| Cgk clearly smaller than Cg | zero point, setting master, wrong reference value | adjust, check the reference value, calibrate the master |
| EV larger than AV | the gage or fixture itself varies | check fixture, clamping, probing, resolution |
| AV larger than EV | appraisers measure differently | sharpen the instruction, fix the measuring spot, train |
| Significant interaction | appraisers measure certain parts differently | clarify form error and measuring position |
| ndc too small | parts hardly vary or the system is too coarse | widen the part selection, improve the system |
| Stability chart violated | drift, wear, temperature change | find the cause, shorten the interval |
| Kappa too small, misses | unclear criterion, lighting, no boundary samples | provide boundary samples, sharpen the criterion |
| QMP too large (VDA 5) | largest contribution in the budget | tackle exactly that contribution |
A case step by step
How do you read the indices in practice? Take the gage R&R from MSA 2: %GRR = 11.1 %, conditionally acceptable. Suppose the customer requires 10 %.
- Check the study. Tolerance 100 µm is correct, resolution 1 µm is 1 % of the tolerance, the ten parts cover a wide range, ndc = 7. The study is fine.
- Compare EV and AV. EV = 1.19 µm, AV = 1.41 µm. AV is slightly larger, so the appraisers contribute at least as much as the gage.
- Look at the pattern. In the appraiser chart of MSA 2 the lines run in parallel: appraiser P2 measures all parts slightly larger. That is a fixed difference, not an interaction. Typical cause: different measuring force.
- Choose a measure. Add to the measuring instruction (“ratchet to the third click, measuring point 5 mm from the shoulder”) and briefly train all three appraisers together.
- Repeat. New study with the same parts. If AV drops noticeably, %GRR falls below 10 %.
No new gage, no investment. Most improvements to measurement systems look like this.
Levers in the measurement process
Proven levers for optimising a measurement process include: gage and setting masters, measuring, clamping and holding forces, measuring points, fixture and alignment, guidance and wear, settling time, measuring strategy, the mean of repeated measurements, calibration of the measuring chain and the environment: vibrations, dust and oil mist, temperature fluctuations and sunlight.
A quick lever is the mean of several measurements: the repeatability shrinks with √n, the appraiser influence does not. In the example from MSA 2, %GRR drops from 11.1 % to 9.8 % with two measurements per part. This only holds if production also measures and averages each part that often; the study must reflect that in the same way.
Recalculated with the values from MSA 2, EV = 1.19 µm, AV = 1.41 µm and T = 100 µm:
| Measurements per part n | EV² / n in µm² | AV² in µm² | Root of the sum in µm | %GRR |
|---|---|---|---|---|
| 1 | 1.415 | 1.980 | 1.843 | 11.1 % |
| 2 | 0.708 | 1.980 | 1.639 | 9.8 % |
| 3 | 0.472 | 1.980 | 1.566 | 9.4 % |
| 5 | 0.283 | 1.980 | 1.504 | 9.0 % |
You can see nicely how the gain shrinks: from one to two measurements %GRR drops noticeably, from three to five hardly at all. The appraiser influence AV always remains. Here it is even the larger part, and that is exactly why in this example a better measuring instruction is the more effective lever than more measurements.
If it still falls short
If that is not enough, a root cause analysis helps, along with aligning limiting factors with development and production (tolerance, production concept, measuring strategy), using existing FMEA results and documenting the measures. It has to be shown that the specification is still met, agreed with the customer if necessary; this can be a conditional approval. Sometimes a different measurement method is the only solution.
It does not help to massage the result: deleting values without reason, adding parts outside the tolerance so that the total variation grows, switching the reference until it fits, or rounding measurements.
What auditors typically want to see about measurement system analysis, from the equipment list through limits and actions to approval, is summarised in the checklist “Measurement system analysis in an audit” (PDF), ready to tick off.
Help in STATlum
On the results page of every study STATlum gives recommendations with cause and remedy, for example “EV clearly larger than AV” or, for VDA 5, the largest contribution in the budget with the matching action.
Download the free versionFree MSA software: what STATlum Free doesCalculate Cg and Cgk online
Frequently asked questions
What to do with %GRR above 30 %?
First check whether repeatability or reproducibility dominates and act there. If that does not help, change the measurement method or agree a solution with the customer.
May I average several measurements?
Yes, if production measures and averages the same way and the study reflects that. The appraiser influence does not shrink by averaging.
Is a conditional approval allowed?
Many customers and company guidelines, such as Bosch Booklet 10, provide for it if it is documented that the specification is still met, agreed with the customer if necessary.
Sources
- Robert Bosch GmbH: Booklet 10, edition 11.2019, chapter 6 and appendix A
- SPC for Excel: Five Common Mistakes with Gage R&R Studies
- Quality Magazine: Gage R&R, the Key to Reducing Measurement Variation
- fabrico: Gauge R&R
- AIAG: Measurement Systems Analysis, Reference Manual, 4th edition
This article explains the method. It does not replace the guideline that applies to your part or the professional assessment of an actual study.