Measurement system analysis

MSA 2: Gage R&R with %GRR and ndc, with example

TL;DR
  • A gage R&R shows how much of the variation comes from the gage and the appraisers.
  • 2 or 3 appraisers measure 10 production parts 2 or 3 times each; ANOVA separates EV, AV and part variation.
  • %GRR up to 10 % is usually acceptable, up to 30 % conditionally acceptable; ndc should be at least 5.

About 11 min read

Contents

What this is about

Many companies know the situation: the early shift measures a part and classifies it as good, the late shift re-measures the same part and finds it just outside the tolerance. Who is right? Often neither entirely, because every measurement contains some scatter, and every appraiser measures a little differently: they position the gage slightly differently, press a little harder or read the value differently.

The gage R&R (also written gauge R&R), in German-speaking countries also called type 2 study, answers the question how large these differences are. Several appraisers measure the same real production parts several times. Afterwards you know which share of the variation comes from the gage, which from the appraisers and which from the parts themselves.

R&R stands for repeatability and reproducibility. Repeatability means: the same appraiser measures the same part several times, how much do the values differ? Reproducibility means: different appraisers measure the same part, how much do their results differ?

Rule of thumb: the parts are supposed to differ, that is intended. The measurement system should add as little as possible. The gage R&R separates the two.

The terms in plain words

Gage R&R, gauge R&R
Two spellings of the same study of repeatability and reproducibility; in German-speaking countries it is called Verfahren 2.
Appraiser
The person who measures. P1, P2 and P3 in the example. What matters are appraisers who use the gage in daily work, not the best metrologist in the company.
Trial
The same part is measured several times by the same appraiser, three times here.
EV, repeatability
Equipment variation. The scatter when the same appraiser measures the same part several times. It comes mainly from the gage and the fixture.
AV, reproducibility
Appraiser variation. The differences between the appraisers: does one of them measure larger on average than the others?
GRR
The variation of the whole measurement system, EV and AV together.
PV, part variation
How much the parts really differ from each other.
Standard deviation and variance
The standard deviation is the typical distance of the values from the mean. The variance is its square. Variances are used because they can be added: variance of the measurement system plus variance of the parts gives the total variance. Standard deviations cannot be added like that.
%GRR
What percentage of the tolerance the spread of the measurement system (6 · GRR) takes up. Smaller is better.
ndc
Number of distinct categories: into how many groups can the measurement system reliably sort the parts? There should be at least 5.
ANOVA
Analysis of variance. A calculation method that divides the total variation among its causes. It sounds complicated but consists only of means, deviations and squares, as the example shows.

Purpose

The gage R&R shows how much of the observed variation comes from the measurement system: from the gage itself (EV) and from the appraisers (AV). The rest is the real variation of the parts (PV).

Why does this matter? Any part close to a tolerance limit can be misclassified because of the measurement system variation: a good part as scrap, a bad one as good. The larger GRR compared with the tolerance, the wider this uncertain zone at the limits. If the measurement system is small compared with tolerance and part variation, the zone is narrow and good and bad parts can be told apart reliably.

If the appraisers play no role, for example with an automatic measuring station, use MSA 3. It is usually preceded by a type 1 study, which checks the gage alone on a reference standard.

How it works

  1. Choose 10 production parts that cover as much of the tolerance or process variation as possible: some rather small, some rather large, some in the middle. Parts only from the middle make ndc look too small, because then the parts hardly have any differences that could be detected.
  2. Number the parts so that the appraiser cannot see the number while measuring, for example on the underside. Otherwise they remember the last value and unconsciously measure towards it.
  3. Select 2 or 3 appraisers who use the gage in daily work. Each measures every part 2 or 3 times. Shuffle the order of the parts randomly in each round.
  4. Everything as in production: same measuring station, same instruction, same fixture. The study should show the measurement system as it really works.
  5. Calculate: the analysis of variance (ANOVA) splits the variation into parts, appraisers, interaction and repetition. EV, AV, GRR and PV follow from it. The example shows every step.
  6. Assess: common limits are %GRR up to 10 % acceptable, up to 30 % conditionally acceptable, above that unacceptable; ndc at least 5 (AIAG). Stricter customer requirements take precedence.

Worked example, step by step

Three appraisers measure a diameter of 10.000 ± 0.050 mm (tolerance T = 0.100 mm = 100 µm) on 10 parts three times each with a micrometer, resolution 0.001 mm. That gives 10 · 3 · 3 = 90 values. The values are artificial and available below as CSV, so you can recalculate every step.

9.9709.9809.99010.00010.01010.020P1P2P312345678910PartMeasured value in mm
Figure 1: Measurements per part and appraiser. The differences between the parts are much larger than the spread within one part.

The chart already tells a lot. The dots of one part are close together, the parts themselves are far apart, part 4 for example clearly below the others. This is the desired picture: the measurement system scatters little, the parts clearly differ. The calculation turns this into numbers.

Step 1: the idea in three questions

The whole calculation answers three simple questions, each with means:

  • How much do the three trials of one appraiser on one part scatter? This becomes EV.
  • How far apart are the means of the three appraisers? This becomes AV.
  • How far apart are the means of the ten parts? This becomes PV.

An example for the first question: appraiser P1 measures part 1 three times and gets 10.005, 10.006, 10.004 mm. The mean of these three is 10.0050 mm, the values deviate from it by 0.00, 1.00, -1.00 µm. Such deviations exist in all 30 combinations of part and appraiser.

Step 2: sums of squares and mean squares

For each of the three questions the squared deviations are added up; that is the sum of squares SS. For the appraisers it looks like this: the grand mean of all 90 values is 10.00356 mm. The appraiser means deviate from it as follows:

AppraiserDeviation from grand mean in µmSquare in µm²
P1-0.390.151
P21.582.489
P3-1.191.413
SSappr=n·r·∑=10·3·4.054µm2=121.6µm2
Each appraiser mean stands for 30 measurements, hence the factor 10 · 3.

A sum of squares grows the more values are involved. To be able to compare the sources, it is divided by the degrees of freedom df, roughly the number of values minus one. For 3 appraisers that is 2. The result is called mean square MS. For the appraisers: 121.6 / 2 = 60.8 µm². The other sources are calculated the same way:

SourceSS in µm²dfMS = SS / df in µm²
Parts7872.29874.7
Appraisers121.6260.8
Appraiser × part interaction16.38180.91
Repetition94.00601.57

New here is the interaction. It asks whether an appraiser measures only certain parts differently, for example only the large ones. Whether it is really present is checked by an F-test: it compares MSint with MSrep. Here MSint is even smaller; the p-value is 0.90. If p is greater than 0.25, the interaction is regarded as absent. It is then added to the repetition:

MSrep′=SSint+SSrepdfint+dfrep=16.38+94.0018+60=1.415µm2
pooled repetition variance

Step 3: from the mean squares to EV, AV and PV

EV is simply the square root of the repetition variance:

EV=MSrep′=1.415µm2=1.19µm
repeatability

For AV there is a small trick. The appraiser means differ not only because the appraisers measure differently, but also because of the chance of repetition. This share is subtracted. And because each appraiser mean consists of 10 · 3 = 30 measurements, it is divided by 30:

AV=MSappr−MSrep′n·r=60.8−1.4230=1.980=1.41µm
reproducibility

The same for the parts. Each part mean consists of 3 · 3 = 9 measurements:

PV=MSparts−MSrep′a·r=874.7−1.429=9.85µm
part variation

EV and AV are added as variances, i.e. squared, and then the square root is taken again:

GRR=EV2+AV2=1.192+1.412=1.415+1.980=1.84µm
measurement system variation
Note that variances add up, standard deviations do not: 1.19 + 1.41 would be 2.60 µm, the correct value is 1.84 µm.

Step 4: %GRR and ndc

Now the comparison with the tolerance. As in the type 1 study, the spread 6 · GRR, which contains practically all deviations of the measurement system, is compared with the tolerance of 100 µm:

%GRR=6·GRRT=6·1.84µm100µm=11.06100=11.1%
The measurement system takes up just over a tenth of the tolerance.

Sometimes GRR is related not to the tolerance but to the total variation TV = √(GRR² + PV²) = 10.02 µm. Then it is 1.84 / 10.02 = 18.4 %. This number says how well the measurement system can tell the parts apart from each other, not how well it fits the tolerance.

ndc=1.41·PVGRR=1.41·9.851.84=7.54→7
number of distinct categories, rounded down

In plain terms: the measurement system can reliably sort the ten parts into 7 groups from “small” to “large”. With only 1 or 2 groups it could merely say “good” or “bad”, which would be too little for process control.

0 %20 %40 %60 %80 %100 %EV7.1 %AV8.4 %GRR11.1 %PV59.1 %10 %30 %6 σ spread in % of tolerance
Figure 2: Variance components as 6 σ in percent of the tolerance, with the common limits of 10 % and 30 %.
InteractiveTry it: %GRR and ndc

This chart needs JavaScript.

Start values as in the example. A smaller part variation worsens ndc and %GRR related to the total variation, but not %GRR related to the tolerance.

What the result means

At 11.1 % the measurement system is just above 10 % and is therefore regarded as conditionally acceptable. ndc = 7 is sufficient. In words: the measurement system is usable, but not so good that it should not be thought about. Whether “conditionally acceptable” is enough depends on the characteristic and on what the customer requires.

Where to start? EV (1.19 µm) and AV (1.41 µm) are of similar size, so both contribute about equally. Figure 3 shows where AV comes from: appraiser P2 measures all parts uniformly slightly larger than P3. Such a fixed difference points to handling, for example how firmly the micrometer is tightened. A more precise instruction, for example “ratchet to the third click”, and brief training often lower AV considerably.

-2-10+1+2P1: mean -0.39 µmP2: mean 1.58 µmP3: mean -1.19 µm12345678910PartDeviation from part mean in µm
Figure 3: Deviation of each appraiser from the part mean. Parallel lines mean a fixed appraiser difference, crossing lines an interaction.
InteractiveTry it: appraiser difference and interaction

This chart needs JavaScript.

A fixed appraiser difference shifts the lines in parallel. An interaction makes them cross: one appraiser then measures only certain parts differently.
ResultMeaningWhat to do
%GRR ≤ 10 %, ndc ≥ 5measurement system acceptablerelease
%GRR 10 to 30 %conditionally acceptabledecide depending on characteristic and customer, plan improvements
%GRR > 30 %not acceptablefind the cause: EV large → gage, fixture; AV large → appraisers, instruction
ndc < 5parts hardly distinguishablecheck the part selection, otherwise improve the measurement system

More in Measurement system not capable.

Common mistakes

  • Parts only from the middle of the tolerance. Then PV is small, ndc drops, and %GRR related to the total variation becomes poor although the measurement system is fine.
  • Appraisers can see the part number or their previous values. They then unconsciously measure towards the known value, and EV appears smaller than it really is.
  • Only the best appraisers measure. The study should show everyday work. If others will measure later, they belong in the study.
  • Fixed order. If parts 1 to 10 are always measured in sequence, warming up or fatigue always affects the same parts.
  • Adding standard deviations instead of variances. GRR = EV + AV is wrong and makes the measurement system look worse than it is.
  • Confusing %GRR related to tolerance and to total variation. Both numbers are correct but answer different questions. Always state in the report which one is meant.

In STATlum

Create an “MSA 2” study on the characteristic and choose the design, the method (ANOVA, average and range, or range) and the number of parts, appraisers and trials. The evaluation shows all indices of this example.

STATlum: configuration of an MSA 2 study
Configuration of an MSA 2 study: rule set, design, method and study size.
STATlum: results of a gage R&R study with %GRR and ndc
Results: %GRR, ndc, variation, smallest tolerance and recommendations.

Frequently asked questions

How many parts, appraisers and trials does a gage R&R need?

Usually 10 parts, 2 or 3 appraisers and 2 or 3 trials. More parts usually improve the estimate more than more trials.

What does %GRR mean?

The share of the measurement system variation in the tolerance or in the total variation. Common: up to 10 % acceptable, up to 30 % conditionally acceptable.

What is ndc?

The number of distinct categories: how many classes the measurement system reliably separates within the part variation. Usually at least 5 are required.

ANOVA or average and range method?

ANOVA uses the data more completely and detects an appraiser × part interaction. The range method is simpler and usually gives similar values when there is no interaction.

What are typical acceptance criteria for %GRR and ndc?

Commonly %GRR up to 10 % is acceptable and up to 30 % conditionally acceptable, with ndc of at least 5. Requirements of your customer or plant take precedence.

Type 1 study vs gage R&R: what is the difference?

A type 1 study checks the gage alone on a reference standard. The gage R&R measures production parts with several appraisers and separates repeatability, reproducibility and part variation. If appraisers play no role, for example with an automatic gage, a type 3 study is used.

Sources

This article explains the method. It does not replace the guideline that applies to your part or the professional assessment of an actual study.