MSA 2: Gage R&R with %GRR and ndc, with example
- A gage R&R shows how much of the variation comes from the gage and the appraisers.
- 2 or 3 appraisers measure 10 production parts 2 or 3 times each; ANOVA separates EV, AV and part variation.
- %GRR up to 10 % is usually acceptable, up to 30 % conditionally acceptable; ndc should be at least 5.
Contents
What this is about
Many companies know the situation: the early shift measures a part and classifies it as good, the late shift re-measures the same part and finds it just outside the tolerance. Who is right? Often neither entirely, because every measurement contains some scatter, and every appraiser measures a little differently: they position the gage slightly differently, press a little harder or read the value differently.
The gage R&R (also written gauge R&R), in German-speaking countries also called type 2 study, answers the question how large these differences are. Several appraisers measure the same real production parts several times. Afterwards you know which share of the variation comes from the gage, which from the appraisers and which from the parts themselves.
R&R stands for repeatability and reproducibility. Repeatability means: the same appraiser measures the same part several times, how much do the values differ? Reproducibility means: different appraisers measure the same part, how much do their results differ?
The terms in plain words
- Gage R&R, gauge R&R
- Two spellings of the same study of repeatability and reproducibility; in German-speaking countries it is called Verfahren 2.
- Appraiser
- The person who measures. P1, P2 and P3 in the example. What matters are appraisers who use the gage in daily work, not the best metrologist in the company.
- Trial
- The same part is measured several times by the same appraiser, three times here.
- EV, repeatability
- Equipment variation. The scatter when the same appraiser measures the same part several times. It comes mainly from the gage and the fixture.
- AV, reproducibility
- Appraiser variation. The differences between the appraisers: does one of them measure larger on average than the others?
- GRR
- The variation of the whole measurement system, EV and AV together.
- PV, part variation
- How much the parts really differ from each other.
- Standard deviation and variance
- The standard deviation is the typical distance of the values from the mean. The variance is its square. Variances are used because they can be added: variance of the measurement system plus variance of the parts gives the total variance. Standard deviations cannot be added like that.
- %GRR
- What percentage of the tolerance the spread of the measurement system (6 · GRR) takes up. Smaller is better.
- ndc
- Number of distinct categories: into how many groups can the measurement system reliably sort the parts? There should be at least 5.
- ANOVA
- Analysis of variance. A calculation method that divides the total variation among its causes. It sounds complicated but consists only of means, deviations and squares, as the example shows.
Purpose
The gage R&R shows how much of the observed variation comes from the measurement system: from the gage itself (EV) and from the appraisers (AV). The rest is the real variation of the parts (PV).
Why does this matter? Any part close to a tolerance limit can be misclassified because of the measurement system variation: a good part as scrap, a bad one as good. The larger GRR compared with the tolerance, the wider this uncertain zone at the limits. If the measurement system is small compared with tolerance and part variation, the zone is narrow and good and bad parts can be told apart reliably.
If the appraisers play no role, for example with an automatic measuring station, use MSA 3. It is usually preceded by a type 1 study, which checks the gage alone on a reference standard.
How it works
- Choose 10 production parts that cover as much of the tolerance or process variation as possible: some rather small, some rather large, some in the middle. Parts only from the middle make ndc look too small, because then the parts hardly have any differences that could be detected.
- Number the parts so that the appraiser cannot see the number while measuring, for example on the underside. Otherwise they remember the last value and unconsciously measure towards it.
- Select 2 or 3 appraisers who use the gage in daily work. Each measures every part 2 or 3 times. Shuffle the order of the parts randomly in each round.
- Everything as in production: same measuring station, same instruction, same fixture. The study should show the measurement system as it really works.
- Calculate: the analysis of variance (ANOVA) splits the variation into parts, appraisers, interaction and repetition. EV, AV, GRR and PV follow from it. The example shows every step.
- Assess: common limits are %GRR up to 10 % acceptable, up to 30 % conditionally acceptable, above that unacceptable; ndc at least 5 (AIAG). Stricter customer requirements take precedence.
Worked example, step by step
Three appraisers measure a diameter of 10.000 ± 0.050 mm (tolerance T = 0.100 mm = 100 µm) on 10 parts three times each with a micrometer, resolution 0.001 mm. That gives 10 · 3 · 3 = 90 values. The values are artificial and available below as CSV, so you can recalculate every step.
The chart already tells a lot. The dots of one part are close together, the parts themselves are far apart, part 4 for example clearly below the others. This is the desired picture: the measurement system scatters little, the parts clearly differ. The calculation turns this into numbers.
Step 1: the idea in three questions
The whole calculation answers three simple questions, each with means:
- How much do the three trials of one appraiser on one part scatter? This becomes EV.
- How far apart are the means of the three appraisers? This becomes AV.
- How far apart are the means of the ten parts? This becomes PV.
An example for the first question: appraiser P1 measures part 1 three times and gets 10.005, 10.006, 10.004 mm. The mean of these three is 10.0050 mm, the values deviate from it by 0.00, 1.00, -1.00 µm. Such deviations exist in all 30 combinations of part and appraiser.
Step 2: sums of squares and mean squares
For each of the three questions the squared deviations are added up; that is the sum of squares SS. For the appraisers it looks like this: the grand mean of all 90 values is 10.00356 mm. The appraiser means deviate from it as follows:
| Appraiser | Deviation from grand mean in µm | Square in µm² |
|---|---|---|
| P1 | -0.39 | 0.151 |
| P2 | 1.58 | 2.489 |
| P3 | -1.19 | 1.413 |
A sum of squares grows the more values are involved. To be able to compare the sources, it is divided by the degrees of freedom df, roughly the number of values minus one. For 3 appraisers that is 2. The result is called mean square MS. For the appraisers: 121.6 / 2 = 60.8 µm². The other sources are calculated the same way:
| Source | SS in µm² | df | MS = SS / df in µm² |
|---|---|---|---|
| Parts | 7872.2 | 9 | 874.7 |
| Appraisers | 121.6 | 2 | 60.8 |
| Appraiser × part interaction | 16.38 | 18 | 0.91 |
| Repetition | 94.00 | 60 | 1.57 |
New here is the interaction. It asks whether an appraiser measures only certain parts differently, for example only the large ones. Whether it is really present is checked by an F-test: it compares MSint with MSrep. Here MSint is even smaller; the p-value is 0.90. If p is greater than 0.25, the interaction is regarded as absent. It is then added to the repetition:
Step 3: from the mean squares to EV, AV and PV
EV is simply the square root of the repetition variance:
For AV there is a small trick. The appraiser means differ not only because the appraisers measure differently, but also because of the chance of repetition. This share is subtracted. And because each appraiser mean consists of 10 · 3 = 30 measurements, it is divided by 30:
The same for the parts. Each part mean consists of 3 · 3 = 9 measurements:
EV and AV are added as variances, i.e. squared, and then the square root is taken again:
Step 4: %GRR and ndc
Now the comparison with the tolerance. As in the type 1 study, the spread 6 · GRR, which contains practically all deviations of the measurement system, is compared with the tolerance of 100 µm:
Sometimes GRR is related not to the tolerance but to the total variation TV = √(GRR² + PV²) = 10.02 µm. Then it is 1.84 / 10.02 = 18.4 %. This number says how well the measurement system can tell the parts apart from each other, not how well it fits the tolerance.
In plain terms: the measurement system can reliably sort the ten parts into 7 groups from “small” to “large”. With only 1 or 2 groups it could merely say “good” or “bad”, which would be too little for process control.
What the result means
At 11.1 % the measurement system is just above 10 % and is therefore regarded as conditionally acceptable. ndc = 7 is sufficient. In words: the measurement system is usable, but not so good that it should not be thought about. Whether “conditionally acceptable” is enough depends on the characteristic and on what the customer requires.
Where to start? EV (1.19 µm) and AV (1.41 µm) are of similar size, so both contribute about equally. Figure 3 shows where AV comes from: appraiser P2 measures all parts uniformly slightly larger than P3. Such a fixed difference points to handling, for example how firmly the micrometer is tightened. A more precise instruction, for example “ratchet to the third click”, and brief training often lower AV considerably.
| Result | Meaning | What to do |
|---|---|---|
| %GRR ≤ 10 %, ndc ≥ 5 | measurement system acceptable | release |
| %GRR 10 to 30 % | conditionally acceptable | decide depending on characteristic and customer, plan improvements |
| %GRR > 30 % | not acceptable | find the cause: EV large → gage, fixture; AV large → appraisers, instruction |
| ndc < 5 | parts hardly distinguishable | check the part selection, otherwise improve the measurement system |
More in Measurement system not capable.
Common mistakes
- Parts only from the middle of the tolerance. Then PV is small, ndc drops, and %GRR related to the total variation becomes poor although the measurement system is fine.
- Appraisers can see the part number or their previous values. They then unconsciously measure towards the known value, and EV appears smaller than it really is.
- Only the best appraisers measure. The study should show everyday work. If others will measure later, they belong in the study.
- Fixed order. If parts 1 to 10 are always measured in sequence, warming up or fatigue always affects the same parts.
- Adding standard deviations instead of variances. GRR = EV + AV is wrong and makes the measurement system look worse than it is.
- Confusing %GRR related to tolerance and to total variation. Both numbers are correct but answer different questions. Always state in the report which one is meant.
In STATlum
Create an “MSA 2” study on the characteristic and choose the design, the method (ANOVA, average and range, or range) and the number of parts, appraisers and trials. The evaluation shows all indices of this example.


Download the free versionExample data as CSVFree MSA software: what STATlum Free doesCalculate a gage R&R onlineFree MSA Excel templates
Frequently asked questions
How many parts, appraisers and trials does a gage R&R need?
Usually 10 parts, 2 or 3 appraisers and 2 or 3 trials. More parts usually improve the estimate more than more trials.
What does %GRR mean?
The share of the measurement system variation in the tolerance or in the total variation. Common: up to 10 % acceptable, up to 30 % conditionally acceptable.
What is ndc?
The number of distinct categories: how many classes the measurement system reliably separates within the part variation. Usually at least 5 are required.
ANOVA or average and range method?
ANOVA uses the data more completely and detects an appraiser × part interaction. The range method is simpler and usually gives similar values when there is no interaction.
What are typical acceptance criteria for %GRR and ndc?
Commonly %GRR up to 10 % is acceptable and up to 30 % conditionally acceptable, with ndc of at least 5. Requirements of your customer or plant take precedence.
Type 1 study vs gage R&R: what is the difference?
A type 1 study checks the gage alone on a reference standard. The gage R&R measures production parts with several appraisers and separates repeatability, reproducibility and part variation. If appraisers play no role, for example with an automatic gage, a type 3 study is used.
Sources
- AIAG: Measurement Systems Analysis, Reference Manual, 4th edition
- NIST/SEMATECH e-Handbook: What are the important issues in a gage study?
- SPC for Excel: ANOVA Gage R&R, part 3 (interaction)
This article explains the method. It does not replace the guideline that applies to your part or the professional assessment of an actual study.