Measurement system analysis

MSA 3: Gage R&R without appraiser influence, with example

TL;DR
  • MSA 3 is the gage R&R for automatic gages and fixtures without appraiser influence.
  • 25 production parts are measured twice each, with real reloading; the measurement system variation is then GRR = EV.
  • Limits as in MSA 2: %GRR up to 10 % acceptable, up to 30 % conditionally acceptable, ndc at least 5.

About 7 min read

Contents

What this is about

Not every measurement system has an appraiser who can influence the result. An automatic gage on the line, a CMM with a fixed program or a fixture into which the part is always loaded the same way: here the machine measures, the person at most presses start. Even such systems scatter. The part never sits exactly the same in the fixture, the temperature varies, the probes have play.

MSA 3, a gage R&R (also written gauge R&R) without appraiser influence, is intended for exactly these cases. It is the little sister of MSA 2: the same idea, just without the question about appraisers. Many production parts are measured twice, and you look at how well the two measurements of each part agree.

Rule of thumb: if the machine measures a part twice, both values should be almost the same. The parts, on the other hand, may differ clearly from each other.

The terms in plain words

Type 3 gage study
A gage R&R without appraisers, for example for automatic gages; also written gauge study.
Without appraiser influence
The result does not depend on who measures. But if the operator loads the part by hand or selects the program, they may have an influence after all; then MSA 2 is the safe choice.
EV, repeatability
The scatter when the same part is measured several times. Without appraisers this is the entire measurement system variation: GRR = EV.
PV, part variation
How much the production parts really differ.
Range R
The difference between the largest and smallest value, here simply the difference between the two measurements of a part.
%GRR and ndc
As in MSA 2: share of the tolerance taken up by the measurement system, and the number of groups into which it can reliably sort the parts.

Purpose

MSA 3 is the gage R&R without appraiser influence: for automatic gages, inline measurement, CMMs running a program or fixtures with a fixed sequence. It shows whether the repeatability of the system is small enough to tell parts apart reliably and check them against the tolerance.

Unlike the type 1 study, MSA 3 uses real production parts, not a standard. This also reveals influences that occur only with real parts, such as form errors, burrs or different surfaces. What MSA 3 does not show is bias: if the automatic gage reads 5 µm too high on all parts, you do not notice it here. That is the job of the type 1 study and the linearity study.

How it works

  1. Choose 25 production parts that cover the variation of production. As in MSA 2: parts only from the middle make ndc look too small.
  2. First run: measure all 25 parts one after the other.
  3. Second run: measure all 25 parts again. So all other parts lie in between. Remove and reload or feed the part every time; with several clamping positions, alternate them. That way the study contains everything that also scatters in production.
  4. Calculate: the analysis of variance with the factor part gives EV and PV. With two measurements it is also easier via the differences. The example shows both ways.
  5. Assess as in MSA 2: common limits are %GRR up to 10 % acceptable, up to 30 % conditionally acceptable, ndc at least 5.

Worked example, step by step

An automatic gage checks a diameter of 10.000 ± 0.050 mm, so the tolerance is T = 0.100 mm = 100 µm. 25 production parts are measured in two runs. The data are artificial and available below as CSV.

9.979.979.989.989.999.9910.0010.0010.0110.0110.0210.02Points on the diagonal:both measurements equalDistance to the diagonal:repeatabilityPosition along the diagonal:part variationMeasurement 1 in mmMeasurement 2 in mm
Figure 1: Measurement 1 against measurement 2. The parts spread widely along the diagonal, the distances to the diagonal are small.

How to read the chart: each dot is a part. The first measurement is on the horizontal axis, the second on the vertical axis. If both measurements were always exactly equal, all dots would lie exactly on the diagonal. So the distance from the diagonal shows the repeatability. How widely the dots are spread along the diagonal shows the part variation. Here the dots are close to the diagonal and far apart: a good sign.

Step 1: differences per part

For each part, take the difference between the two measurements. The first six parts, differences in µm:

PartMeas. 1 in mmMeas. 2 in mmDifference dd²|d|
110.0009.998242
210.00410.004000
39.9969.994242
49.9869.986000
59.9929.994-242
69.9849.985-111
………………
Sum of all 255832

Step 2: repeatability EV

With exactly two measurements per part there is a simple formula: the repeatability variance is the sum of the squared differences divided by 2 · n. That is exactly what the analysis of variance also gives, just without the detour.

MSrep=∑d22·n=58µm22·25=1.160µm2
repeatability variance
EV=MSrep=1.160=1.08µm
repeatability, at the same time GRR

Why divide by 2 · n? Each difference contains the scatter of two measurements, so d² is on average twice as large as the variance of a single measurement.

Step 3: part variation PV

For the part variation, take the mean of the two measurements of each part. Their grand mean is 9.99428 mm. The squared deviations of the part means from the grand mean give the mean square of the parts, MSparts = 287.3 µm². It still contains a small share of repeatability, which is subtracted. Divided by 2, because each part mean consists of two measurements:

PV=MSparts−MSrepr=287.3−1.162=11.96µm
part variation

Step 4: cross-check with the ranges

An older, even simpler way uses the absolute differences instead of the squares. Their mean is called R̄: 32 µm / 25 = 1.28 µm. A statistical conversion factor (d₂ = 1.128 for two measurements) turns it into a standard deviation:

EV≈R̄1.128=1.28µm1.128=1.13µm
close to the result of the analysis of variance

The ranges also help to find faulty measurements. If a difference lies above the limit 3.267 · R̄, something probably went wrong with this part, for example dirt in the fixture. Then clarify the cause before using the result.

01234135791113151719212325average range 1.28 µmupper control limit 4.18 µmPartRange in µm
Figure 2: Range of the two measurements per part. None lies above the upper control limit 3.267 · R̄ = 4.2 µm, so there is no sign of faulty measurements.

Step 5: %GRR and ndc

%GRR=6·EVT=6·1.08µm100µm=6.46100=6.5%
from the ranges 6.8 %, related to the total variation 9.0 %
ndc=1.41·PVEV=1.41·11.961.08=15.66→15
rounded down
InteractiveTry it: repeatability and part variation

This chart needs JavaScript.

Tolerance 0.100 mm. As the repeatability grows, the points move away from the diagonal and %GRR rises. With a small part variation they move together and ndc drops.
InteractiveTry it: %GRR and ndc

This chart needs JavaScript.

Without appraisers GRR equals EV. Start values as in the example.

What the result means

With %GRR = 6.5 % the automatic gage takes up only a small part of the tolerance. It is acceptable for this tolerance, ndc = 15 is well above 5. In words: if the gage measures a part twice, the values differ by about 1.3 µm. The parts differ from each other by many times that amount.

This says nothing about bias, as described above. If %GRR were too high, the causes would usually be fixture, probing, temperature or program. More in Measurement system not capable.

Common mistakes

  • Part not removed between measurements. If it stays clamped, the scatter from loading is missing and the result is too good.
  • Both measurements directly one after the other. Then the gage measures twice under exactly the same conditions. First all parts, then all again.
  • MSA 3 although the operator has an influence. If someone loads the part by hand and can tilt it, they belong in the study, i.e. MSA 2.
  • Only parts from the middle of the tolerance. As in MSA 2, ndc is then artificially small.
  • Ignoring conspicuous ranges. A single large difference points to a disturbance, not to poor repeatability.

In STATlum

Create an “MSA 3” study on the characteristic and choose the number of parts and trials. The measurements form a table with one column per trial, also for import. The evaluation shows all indices of this example.

STATlum: measurement table of an MSA 3 study
Measurements of an MSA 3 study: one row per part, one column per trial.
STATlum: results of an MSA 3 study
Results of an example with too much variation: %GRR above 30 %, with recommendations.

Frequently asked questions

When do I use MSA 3 instead of MSA 2?

When the result does not depend on the appraiser, for example with automatic gages or inline measurement. If the appraiser influences loading or the sequence, MSA 2 is the safe choice.

How many parts does MSA 3 need?

Usually 25 parts with 2 measurements each. More parts mainly improve the estimate of the part variation and thus ndc.

Do other limits apply to MSA 3?

No, the usual limits are the same as for MSA 2: %GRR up to 10 % acceptable, up to 30 % conditionally acceptable, ndc at least 5. Your customer’s requirement is what counts.

Sources

This article explains the method. It does not replace the guideline that applies to your part or the professional assessment of an actual study.