Measurement system analysis

Gage resolution: why 5 % of the tolerance?

A display that is too coarse makes a gage look better than it is. Where the 5 % rule comes from and what rounding does to spread and Cg.

About 5 min read

In brief
  • Check the resolution first: AIAG requires at most 10 %, VDA 5 at most 5 % of the tolerance.
  • The less the gage varies, the finer the display must be.
  • Too coarse a resolution shrinks s and inflates Cg, seemingly better.
  • Never round values afterwards.
Contents

What this is about

With a ruler that only has centimetre marks you cannot measure the thickness of a sheet of paper. You can try, but you will always read “0 cm”, or “1 cm” if you are generous. The ruler is not broken; it is simply too coarse for this task.

With gages in production the problem is the same, only less obvious. The resolution is the smallest step a gage can display: often 0.01 mm for a caliper, 0.001 mm for a micrometer. If this step is too coarse compared with the tolerance, you do not see the differences between the parts, and all indices of a measurement system analysis become unreliable, often too good.

Rule of thumb: resolution is the first hurdle. Only when it is fine enough is it worth calculating Cg, Cgk or %GRR.

The rule

Before assessing a gage, check its resolution, the smallest step of the display. The AIAG MSA manual gives the rule of tens: the resolution should be at most one tenth of the tolerance or the process variation; whether the gage tells the parts apart well enough is shown there in addition by ndc ≥ 5 from the gage R&R. VDA 5 is stricter and requires at most 5 % of the tolerance, and additionally counts the resolution as an uncertainty contribution uRE = RE / (2 √3). Many company guidelines, such as Bosch Booklet 10, adopt the 5 %.

%RE=RET·100%
resolution relative to the tolerance

A worked example: the tolerance is 25.000 ± 0.030 mm, i.e. T = 0.060 mm = 60 µm. Then 10 % of the tolerance is 6 µm and 5 % is 3 µm. Which instruments qualify?

InstrumentResolution%RERule of tens (10 %)5 % rule
Caliper0.01 mm = 10 µm10 / 60 = 16.7 %not metnot met
Gage with 5 µm steps0.005 mm = 5 µm5 / 60 = 8.3 %metnot met
Micrometer0.001 mm = 1 µm1 / 60 = 1.7 %metmet

So for this characteristic only the micrometer qualifies. The caliper would already be too coarse for the rule of tens.

Why a coarse resolution misleads

A coarse display rounds the real variation away. If it is of the same order as the spread, almost all measurements show the same value, and the standard deviation becomes far too small, in the extreme zero. Then Cg and Cgk look excellent although the gage cannot tell parts apart at all. Wheeler calls such data “chunky”: control charts and indices lose their meaning.

Example: the same measurements, rounded differently

A standard of 25.000 mm is measured 50 times very finely (0.1 µm), tolerance 25.000 ± 0.030 mm. The same values are then rounded to coarser resolutions, as if a different instrument had been used. The data are artificial and available as CSV.

What does “rounding to a resolution” mean? You take the nearest step. The first three values 25.0008, 25.0023, 24.9995 mm become 25.001, 25.002, 25.000 mm at 1 µm, 25.000, 25.000, 25.000 mm at 5 µm and 25.00, 25.00, 25.00 mm at 10 µm. So the differences between the values disappear step by step.

0.1 µm1 µm5 µm10 µmReference value24.99625.00025.00425.008The same 50 measurements, rounded to the resolution (left)
Figure 1: At 5 and 10 µm almost all values fall onto a few steps.
Resolution%REdistinct valuessCgCgk
0.1 µm0.2 %351.25 µm1.601.50
1.0 µm1.7 %61.26 µm1.591.49
2.0 µm3.3 %41.38 µm1.451.36
5.0 µm8.3 %20.71 µm2.832.78
10.0 µm16.7 %10.00 µmnot computablenot computable

Up to 2 µm s stays close to the true value (1.25 µm). At 5 µm (8.3 % of the tolerance) the gage shows only 2 distinct values, s drops to 0.71 µm and Cg rises to 2.83. At 10 µm all 50 measurements show the same value, s is zero and Cg can no longer be computed: a seemingly perfect gage that in truth resolves nothing. If the mean lies between two steps instead, the values jump back and forth and s becomes too large; that distorts too.

InteractiveTry it: change the resolution

This chart needs JavaScript.

The same 50 measurements as in the example, start value 1 µm. Light blue the fine values, strong the rounded ones. Watch when s collapses and Cg shoots up.

What this means in practice

  • Always check the resolution first; with too coarse a resolution Cg, Cgk, %GRR and ndc are meaningless.
  • Many identical values in a series are a warning sign, even when the 5 % is met.
  • The display alone does not count: an instrument with a fine display but a large spread of its own still does not resolve more finely.
  • Do not round values afterwards, not even in spreadsheets; it has the same effect as a coarse display.

In STATlum %RE appears in the results of MSA 1, and the resolution as contribution uRE in the VDA 5 budget.

Common mistakes

  • Confusing resolution with accuracy. A display with four decimal places says nothing about whether the last digit is correct.
  • Checking the resolution against the wrong quantity. What counts is the tolerance of the characteristic (or the process variation), not the measuring range of the instrument.
  • Rounding values in Excel. A neat table with two decimal places turns a micrometer into a caliper.
  • Trusting good indices despite many identical values. If 40 of 50 measurements show the same value, s is too small and Cg too large.

Frequently asked questions

Is a resolution of 10 % of the tolerance enough?

Just about, according to the rule of tens of the AIAG manual; VDA 5 and many company guidelines require at most 5 %. The example shows that even a resolution below 10 % can distort the spread considerably when the gage itself varies little.

What is the difference between resolution and accuracy?

Resolution is the smallest display step. How accurately the gage measures also depends on repeatability, bias and calibration.

May I round measurements for the evaluation?

No. Rounded values act like a coarser resolution and distort the spread and the indices.

Sources

This article explains the method. It does not replace the guideline that applies to your part or the professional assessment of an actual study.