Measurement system analysis

MSA 4: Measurement system stability study with control chart and example

TL;DR
  • A stability study shows whether a measurement system measures the same over time.
  • A stability part is measured 3 to 5 times at regular intervals and plotted on a control chart with limits x̿ ± A₂R̄.
  • It is unstable with a value outside, 9 values on one side or 6 values rising or falling.

About 8 min read

Contents

What this is about

A gage that measures well today does not necessarily do so in three months. Probes wear, inserts are replaced, someone knocks against the fixture, the hall is warmer in summer than in winter. Such changes often creep in, and nobody notices because every single measurement still looks plausible.

The stability study, called MSA 4 here, catches such changes. You regularly measure the same part, for example every morning, and enter the results in a control chart. As long as the measurement system stays the same, the points fluctuate randomly around a centre line. If something changes, the pattern stands out: points jump out, run on one side or rise steadily.

Rule of thumb: the type 1 study and gage R&R are photos, the stability study is a film.

The terms in plain words

Stability part
The part that is measured again and again: a reference standard or a marked production part that does not change itself. It stays at the measuring station and is kept clean and at the same temperature.
Subgroup
The measurements at one time point, 3 measurements every morning in the example.
Control chart
A diagram with time to the right and the measured value upwards, plus a centre line and two limits. The limits show how far chance alone lets the values scatter.
Mean chart (x̄ chart)
Shows the mean of each subgroup. It reacts to shifts in location.
Range chart (R chart)
Shows the difference between the largest and smallest value of each subgroup. It reacts to increasing spread.
Control limits
The upper and lower limit of the chart (UCL and LCL). They are placed so that a stable system almost never goes beyond them, only about 3 in 1000 points.
Nelson rules
Patterns that are very unlikely with pure chance. If one occurs, something has probably changed.

Purpose

A stability study shows whether a measurement system measures the same over time. The type 1 study and gage R&R are snapshots; MSA 4 reveals drift, jumps and wear before wrong values reach production. The AIAG MSA manual (4th edition) calls it stability.

The benefit is very practical: if a change is detected early, only the parts since the last unremarkable measurement need to be rechecked, not a whole week of production. And a calm chart over months is a good reason to extend the calibration interval.

How it works

  1. Choose a stability part: a reference standard or a marked production part from the middle of the measuring range. Determine its reference value in the metrology lab.
  2. Measure regularly: 3 to 5 measurements per time point, several times per shift at first. Always measure under the same conditions as in production. Record temperature and special events, such as “probe replaced”.
  3. Calculate after 20 to 25 time points: mean and range chart with limits from the data. Fewer time points give uncertain limits.
  4. Assess with Nelson rules 1 to 3: a value outside the limits, 9 values in a row on one side of the centre line, 6 values in a row rising or falling.
  5. Continue: the limits stay fixed, new points are entered continuously. If the chart stays calm, the interval may become longer, for example weekly instead of daily.

Worked example, step by step

A setting master with reference value 20.002 mm is measured 3 times at the start of the shift on 25 days. The data are artificial (spread 0.8 µm, with a jump of 1.1 µm from day 17 built in to show what the chart finds) and available below as CSV.

Step 1: mean and range per day

For each day calculate the mean of the three measurements and the range, i.e. largest minus smallest value. The first three days:

DayValues in mmMean x̄ in mmRange R in µm
120.0030 / 20.0032 / 20.002120.002771.1
220.0014 / 20.0011 / 20.002020.001500.9
320.0012 / 20.0009 / 20.002220.001431.3
…………

Step 2: centre line and average range

From the 25 daily means, take the mean again; that is the centre line of the chart. Likewise the mean of the 25 ranges.

x̿=∑x̄j25=20.00246mm
centre line of the mean chart
R̄=∑Rj25=1.22µm
average range, centre line of the range chart

Step 3: control limits

How far may the daily means fluctuate randomly around the centre line? That depends on the spread within a day, which is known from R̄. A table factor A₂ converts R̄ directly into the distance of the limits. For 3 measurements per time point A₂ = 1.023. The factors are in every textbook on control charts; behind them is the rule “centre line ± 3 standard deviations of the mean”.

A2·R̄=1.023·1.22µm=1.24µm
distance of the limits from the centre line
LCL=20.00246−0.00124=20.00122mm
UCL=20.00246+0.00124=20.00371mm
control limits of the mean chart
UCLR=D4·R̄=2.574·1.22µm=3.13µm
range chart, the lower limit is zero for 3 measurements

Step 4: draw the chart and check the rules

20.00020.00120.00220.00320.00420.005Centre line x̿Control limits x̿ ± A₂R̄Reference value135791113151719212325Time point (day)Mean in mm
Figure 1: Mean chart. From day 17 the values lie higher, and on day 25 the upper limit is exceeded. Because the centre line is computed from all days, it lies between both levels.

Now go through the rules. Rule 1, a point outside the limits: on day 25 the mean is 20.00397 mm, above the upper limit of 20.00371 mm. Rule 2, 9 points in a row on one side: does not occur here. Rule 3, 6 points in a row rising or falling: does not occur either.

Looking more closely, the jump is visible anyway: days 1 to 16 average 20.00212 mm, days 17 to 25 20.00307 mm, about 1.0 µm higher. The chart only flagged this on the last day. That is typical of small jumps, which is why it pays to look at the pattern with your own eyes as well as applying the rules.

0123upper limit D₄R̄ = 3.13 µm135791113151719212325Time point (day)R in µm
Figure 2: Range chart, unremarkable. The spread per time point has not changed.

The range chart stays calm; no range lies above 3.13 µm. So the spread within a day has stayed the same; only the location has shifted. The bias over the whole period is x̿ − xref = 20.00246 − 20.002 = +0.46 µm.

InteractiveTry it: jump, drift and spread

This chart needs JavaScript.

Start values and points as in the example. A small jump is often detected late because it also moves the centre line; a drift mostly shows up as points outside the limits and long runs on one side.
InteractiveTry it: how fast does the chart detect a shift?

This chart needs JavaScript.

Probability that a single mean lies outside the 3σ limits, depending on the shift and on the sample size n. More measurements per time point make the chart much more sensitive.

What the result means

The measurement system is not stable. The spread is unchanged, but the location has shifted by about 1 µm from day 17. Typical causes of such a pattern are a replaced insert, a newly set zero point or a modified fixture. A look at the notes for days 16 and 17 often helps.

What should be done? Find and fix the cause. Then check whether capability has to be shown again, for example with the type 1 study. And afterwards determine the limits again from a stable period, because the current limits contain the jump. Bosch Booklet 10 instead sets the limits from the tolerance, see Booklet 10.

PatternPossible cause
jump in the meansinsert replaced, zero point reset, fixture changed
slow driftwear, temperature, contamination
ranges increaseplay in the mechanics, loose fixture, new appraiser
pattern following the time of dayroom temperature, sunlight, machine warming up

Common mistakes

  • Changing stability part. If sometimes one part and sometimes another is measured, you see the differences between the parts instead of the change in the measurement system.
  • Limits from the tolerance instead of from the data. The tolerance is much wider than the spread of the measurement system, so small changes remain invisible. This only makes sense if a guideline such as Booklet 10 explicitly requires it.
  • Only one value per time point. Then the range chart is missing, and an increasing spread goes unnoticed.
  • Special events not recorded. Without the note “probe replaced”, the cause of a jump is hard to find later.
  • Limits not recalculated after a change. The old limits then no longer fit the system.

In STATlum

Create an “MSA 4” study on the characteristic and choose the sample size, the source of the limits (default: from the data) and the rules (default: Nelson 1 to 3). Each time point is one row; the evaluation marks rule violations.

STATlum: measurements of an MSA 4 study
Measurements of an MSA 4 study: one row per time point, one column per measurement.
STATlum: results of an MSA 4 study
Results of a stable setting master: no rule violation, location as the reference value.

Frequently asked questions

How often must the stability part be measured?

There is no fixed rule. A short interval at the start, several times per shift, is common and is extended if the chart stays calm. In running production one check at the start of each shift is often enough.

Is there a limit value for stability?

The AIAG manual gives no index. Stable means: no rule violation on the control chart. The size of the bias and of the spread is judged against the tolerance in addition.

Is MSA 4 always required?

No. It is useful above all when stable long-term behaviour is not certain, for example with indices close to the limit, wear, temperature influence or drifting sensors.

What is the difference between stability and linearity?

Stability (MSA 4) shows whether the measurement system changes over time, for example through drift or wear; the same part is always measured. Linearity (MSA 5) shows whether the bias changes across the measuring range; several reference standards are measured for it.

Sources

This article explains the method. It does not replace the guideline that applies to your part or the professional assessment of an actual study.