Bosch Booklet 10: Capability of measurement and test processes explained
- Bosch Booklet 10 (11.2019) combines several procedures into one capability verification for a characteristic.
- First the resolution, then procedure 1, then 2 or 3 depending on appraiser influence; further procedures only when needed.
- The weakest required procedure determines the overall verdict; binding minimum values are set in CDQ 0301.
Contents
What this is about
Anyone who supplies Bosch, or works for a customer who follows Bosch, often receives a brief requirement: “capability verification according to Booklet 10”. That sounds like a single study but is more like a checklist. Booklet 10 says which studies are needed for a characteristic, with which limits they are assessed and how an overall verdict is formed from several results.
The individual studies are old acquaintances: procedure 1 corresponds to MSA 1, procedure 2 to the gage R&R and so on. What is new is mainly the logic around them: first you check the resolution, then the instrument alone, then the measurement process, and in the end the weakest link decides.
The terms in plain words
- Capability verification
- The set of all required studies for a characteristic, with an overall verdict: capable, conditionally capable or not capable.
- Procedures 1 to 7
- The studies of the booklet. Procedures 1 to 5 for measuring inspections, 6 and 7 for pass/fail decisions.
- Appraiser influence
- Whether the person measuring can influence the result. This determines whether procedure 2 or 3 applies.
- %RE
- The resolution as a percentage of the tolerance. The first hurdle: if the display is too coarse, any further verification is pointless.
- Conditionally capable
- The result lies between the limits for “capable” and “not capable”. Then you must improve or justify and document an approval.
- CDQ 0301
- The Bosch directive which, according to the booklet, sets the binding minimum values. The booklet itself explains the procedures.
Purpose
Booklet 10 “Capability of Measurement and Test Processes” from Bosch’s series on technical statistics applies to Bosch and many suppliers. It does not describe a single study but a capability verification for one characteristic: which procedures are required depends on the kind of characteristic and on the measurement process, and the weakest required procedure determines the overall verdict. The current edition is 11.2019; according to the booklet, the Bosch directive CDQ 0301 in its valid version is binding for the minimum values.
The advantage of this structure: no time is wasted. If the resolution is too coarse, procedure 1 is not needed. If procedure 1 fails, procedure 2 is not needed, because an instrument that already scatters too much on the standard will not get better with appraisers and production parts. And because the booklet sets fixed limits and a clear order, verifications from different departments and suppliers are comparable.
How it works
- Check the resolution: below 5 % of the tolerance. Between 5 and 10 %, check whether a finer instrument exists; above 10 % a verification makes no sense.
- Procedure 1 is normally a prerequisite for all others.
- Procedure 2 if the appraiser can have an influence, otherwise procedure 3.
- Procedures 4 and 5 (linearity, stability) only when needed, procedures 6 and 7 for pass/fail inspections.
- Overall verdict: the weakest required procedure decides. The verification applies only to the characteristic studied; it is best to agree beforehand which procedures apply.
The booklet calls them procedures 1 to 7. In STATlum they are called MSA 1 to MSA 7A, numbered differently from procedure 4 on:
| Booklet 10 | Content | STATlum | Criterion per booklet |
|---|---|---|---|
| Procedure 1 | Bias and repeatability on a standard | MSA 1 | Cg, Cgk ≥ 1.33 |
| Procedure 2 | Gage R&R with appraiser influence | MSA 2 | %GRR ≤ 10 % capable, ≤ 30 % conditional; ndc ≥ 5 |
| Procedure 3 | Gage R&R without appraiser influence | MSA 3 | as procedure 2 |
| Procedure 4 | Linearity | MSA 5 | type 1 study per standard or AIAG |
| Procedure 5 | Stability | MSA 4 | stability chart without violation |
| Procedure 6 | Test decisions, signal detection | MSA 7 analytic | %GRR ≤ 10 % capable, ≤ 30 % conditional |
| Procedure 7 | Test decisions, kappa | MSA 6, MSA 7A | κ ≥ 0.9 capable, ≥ 0.7 conditional |
Worked example, step by step
The characteristic is the diameter 10.000 ± 0.050 mm (T = 100 µm) from the articles MSA 2 and VDA 5, measured with a micrometer. The appraiser has an influence and linearity and stability are considered established, so procedures 1 and 2 are required.
Why exactly these two? The appraiser applies the micrometer by hand, so they can influence the result: procedure 2 instead of 3. Linearity and stability would only be studied for a micrometer in a small range if there were a specific reason. The interactive chart shows the decision path.
Step 1: resolution
The micrometer displays 0.001 mm = 1 µm, and the tolerance is 100 µm wide. So the display divides the tolerance into 100 steps, more than enough. It would only get tight below 20 steps (5 %).
Step 2: procedure 1
Procedure 1 uses the 25 measurements on the setting master from the VDA 5 example (Booklet 10 requires at least 25, 50 are usual): x̄ = 10.00076 mm, xm = 10.0003 mm, s = 0.970 µm.
The calculation is as described in MSA 1. 20 % of the tolerance is 20 µm. The spread of the instrument is 6 · s = 6 · 0.970 = 5.82 µm. So it fits into the band a good three times:
For Cgk, half the band of 10 µm is reduced by the bias of 0.46 µm and compared with half the spread 3 · s = 2.91 µm. Both values are far above 1.33: the instrument alone is very good.
Step 3: procedure 2
Procedure 2 takes the gage R&R from MSA 2 with GRR = 1.84 µm and PV = 9.85 µm.
How EV, AV and GRR arise from the 90 values is shown step by step in the article MSA 2. Here only the result counts: the spread of the measurement system takes up just over a tenth of the tolerance.
Step 4: overall verdict
| Procedure | Indices | Verdict per booklet |
|---|---|---|
| Resolution | %RE = 1.0 % | below 5 % |
| Procedure 1 | Cg = 3.44, Cgk = 3.28 | capable |
| Procedure 2 | %GRR = 11.1 %, ndc = 8 | conditionally capable |
| Verification | weakest procedure decides | conditionally capable |
What the result means
The table above shows the typical picture of a verification: one link is very strong, another one is borderline. Only the borderline one counts for the overall verdict.
The instrument itself is very good, the measurement process with appraisers is just above 10 % at 11.1 %. The verification is therefore only conditionally achieved. According to chapter 6 of the booklet, the process must then be improved or a conditional approval justified and documented, see also Measurement system not capable. For comparison: according to VDA 5 the same measurement process is suitable with QMP = 9.6 %.
The difference between the rule sets is not a calculation error. VDA 5 looks at the uncertainty of a measured value, Booklet 10 at the variation of the measurement system with fixed limits per procedure. The customer decides which rule set applies.
Common mistakes
- Skipping the resolution. If it is too coarse, all further results are worthless, no matter how good they look.
- Procedure 3 although the appraiser has an influence. Anyone who loads the part or applies the instrument by hand belongs in the study.
- Reporting only the best result. The overall verdict always follows the weakest required procedure.
- ndc rounded as in AIAG. Booklet 10 rounds to the nearest integer, AIAG rounds down. State in the report which rule applies.
- Limits not checked. The binding minimum values are in CDQ 0301 in its valid version, not necessarily in the booklet.
- Transferring the verification to another characteristic. It applies only to the characteristic studied with its tolerance.
In STATlum
Create a “Bosch Booklet 10” study on the characteristic. It has no measurements of its own but combines the studies of the characteristic: set the type of characteristic, appraiser influence, linearity and stability, and assign a study to each required procedure. With the “Bosch Booklet 10” rule set, the individual studies use the criteria and study size of the booklet.

Download the free versionFree MSA software: what STATlum Free doesCalculate Cg and Cgk online
Frequently asked questions
Do I have to run all seven procedures?
No. For a continuous characteristic procedure 1 plus 2 or 3 is usually required; 4 and 5 only when needed. Pass/fail inspections use procedures 6 and 7.
Which limits apply?
According to Booklet 10 (11.2019) Cg and Cgk at least 1.33, %GRR up to 10 % capable and up to 30 % conditionally capable, kappa from 0.9 capable and from 0.7 conditionally capable. The minimum values of CDQ 0301 in its valid version are binding.
Is Booklet 10 the same as AIAG MSA?
No. Procedure 2 follows AIAG MSA directly, the type 1 study with Cg and Cgk does not exist there, and for attribute inspections Booklet 10 uses Fleiss’ instead of Cohen’s kappa.
What sample size does Booklet 10 require?
For procedure 1 at least 25 measurements on the reference standard, 50 are common. The articles on MSA 2 to MSA 7A describe the sample size of the other procedures.
Sources
- Robert Bosch GmbH: Booklet 10, Capability of Measurement and Test Processes, edition 11.2019
- AIAG: Measurement Systems Analysis, Reference Manual, 4th edition
- Dietrich: Wo liegen die Unterschiede? Vergleich von MSA und VDA Band 5 (German), QZ 6/2011
- symestic: What is a measurement system analysis? (German)
This article explains the method. It does not replace the guideline that applies to your part or the professional assessment of an actual study.