Measurement system analysis

MSA 7: Attribute gage study with signal detection and acceptance curve

TL;DR
  • MSA 7 assesses a limit gage or pass/fail test device on parts with a measured reference value.
  • 50 parts, many close to the limits, 3 appraisers, 3 trials each; d is the width of the zone with inconsistent ratings.
  • %GRR = d / T; according to AIAG below 10 % acceptable, up to 30 % depending on the application.

About 7 min read

Contents

What this is about

A limit gage knows only two answers: the go side passes over the shaft, the no-go side does not, then the part is OK. That is quick, robust and needs no electronics. But what happens with a part that lies exactly on the limit? Sometimes the gage just passes, sometimes not, depending on how it is applied, how warm the part is and who inspects.

So far from the limits the rating is clear; close to a limit it varies. Signal detection, MSA 7 here, measures how wide this uncertainty zone is. This requires parts whose exact size is known, measured with a more accurate instrument. Then you look at from which size the ratings start to vary and from which size they are clear again.

Rule of thumb: the narrower the uncertainty zone compared with the tolerance, the better the gage. Its width corresponds roughly to the 6 · GRR spread of a measuring gage.

The terms in plain words

Limit gage
A gage with a go and a no-go side, for example a ring gage or plug gage. It does not measure; it decides.
Reference value
The exact size of each part, measured with a much more accurate instrument, for example in the metrology lab.
Code + − x
“+”: all 9 ratings OK; “−”: all 9 ratings not OK; “x”: inconsistent, sometimes one, sometimes the other.
Uncertainty zone d
The range around a limit in which the ratings are inconsistent. d₁ at the lower limit, d₂ at the upper.
%GRR
The mean uncertainty zone relative to the tolerance. Comparable with the %GRR of a gage R&R.
Acceptance curve
Shows for each size how often a part is accepted. With a perfect gage it jumps vertically from 0 to 100 % at the limits.

Purpose

Signal detection turns a pass/fail inspection into an index that can be compared with measuring gages: a %GRR as in a gage R&R. The AIAG MSA manual (4th edition) calls it the signal detection approach.

The prerequisite is measured reference values. If there are none, for example in visual inspection, the assessment uses kappa as in MSA 6; in STATlum this variant is called MSA 7 consensus. For automatic gages there is MSA 7A.

How it works

  1. Choose 50 production parts and measure each one accurately; that is the reference value. The values range from slightly below the lower to slightly above the upper limit, with many parts close to both limits, because that is where it matters. A few parts from the middle serve as a check.
  2. 3 appraisers rate every part 3 times with the gage, reshuffled every time, without knowing the reference values.
  3. Code: each part gets “+”, “−” or “x”.
  4. Sort and read off: sorted by reference value, the “x” parts sit at the limits. Their widths d1 and d2 give %GRR.
  5. Assess as in a gage R&R: commonly below 10 % acceptable, up to 30 % depending on the application, above that unacceptable.

Worked example, step by step

A diameter of 10.000 ± 0.030 mm, i.e. tolerance T = 0.060 mm = 60 µm, is checked with a limit gage. The lower limit is 9.970 mm, the upper 10.030 mm. 50 parts with a measured reference value are rated three times by each of three appraisers, 9 ratings per part. The data are artificial and available below as CSV.

LSL 9.970USL 10.030d = 9 µmd = 8 µm9.959.979.9910.0110.0310.05OKnot OKReference value in mm; one column of 9 ratings per part (3 appraisers, 3 trials)
Figure 1: All 450 ratings over the reference value. 8 parts receive inconsistent ratings (bold), all of them at the limits.

Step 1: sort and code the parts

Sort all 50 parts by reference value and write down for each how many of the 9 ratings are OK. Only the region around the limits is of interest. At the lower limit 9.970 mm it looks like this:

Reference value in mmRatings OKCode
9.9640 / 9−
9.9660 / 9−
9.9672 / 9x
9.9694 / 9x
9.9717 / 9x
9.9738 / 9x
9.9759 / 9+
9.9769 / 9+

Up to 9.966 mm all appraisers always say not OK, from 9.975 mm always OK. In between lie the “x” parts: the closer the size gets to the limit, the more ratings flip. The same at the upper limit 10.030 mm, just mirrored:

Reference value in mmRatings OKCode
10.0269 / 9+
10.0288 / 9x
10.0296 / 9x
10.0312 / 9x
10.0322 / 9x
10.0340 / 9−

Step 2: read off the uncertainty zones

At each limit, d is the distance between the last part that is still consistently “−” and the first one that is consistently “+” again. So you take the two parts that frame the “x” region:

d1=9.975−9.966=0.009mm=9µm
uncertainty zone at the lower limit
d2=10.034−10.026=0.008mm=8µm
uncertainty zone at the upper limit

Step 3: %GRR

Take the mean of both zones and relate it to the tolerance. According to AIAG, d corresponds roughly to six times the GRR standard deviation, i.e. the spread that is also used to form %GRR for measuring gages.

d=d1+d22=9+82=8.5µm
mean uncertainty zone
%GRR=dT=8.5µm60µm=14.2%
comparable with the %GRR of a gage R&R
This method is rough: the width depends on how densely the parts lie at the limit. With parts 2 µm apart, d cannot be determined more accurately than about 2 µm. So place many parts close to the limits.
InteractiveTry it: spread of the gage

This chart needs JavaScript.

Start: the ratings of the example (gage spread 1.6 µm, appraiser B +0.6 µm, appraiser C −0.4 µm). With a larger spread the uncertainty zones and thus %GRR grow. An offset moves both zones in the same direction.

Step 4: acceptance curve

Plotting the share of acceptances for each part over the reference value gives the acceptance curve. With a perfect gage it would be a rectangle: 0 % left of the lower limit, 100 % in between, 0 % again on the right. A real gage has sloping edges. The steeper the edges, the sharper the inspection separates. The AIAG analytic method (8 parts with 20 trials each) estimates the repeatability and bias of the gage from this curve.

0 %25 %50 %75 %100 %9.959.979.9910.0110.0310.05Reference value in mmaccepted
Figure 2: Share of accepted ratings per part. Between the edges parts are almost always accepted, outside almost never.
InteractiveTry it: acceptance curve and number of ratings

This chart needs JavaScript.

Solid: probability of acceptance. Dashed: probability of inconsistent ratings for the chosen number of ratings; more ratings reveal a wider uncertainty zone.

What the result means

With %GRR = 14.2 % the gage inspection is conditionally acceptable according to AIAG. In words: parts whose size is less than about 4 µm from a limit, i.e. within half the width of the uncertainty zone, are not reliably sorted correctly. A part of 9.968 mm, i.e. 2 µm below the lower limit, may well pass as good.

Whether this is enough depends on how many parts in production lie that close to the limits. If the process is well centred and narrow, hardly any parts reach the uncertainty zone, and the gage is sufficient. If many parts lie close to a limit, a better or new gage helps, or a measuring gage proven with a type 1 study and gage R&R.

Common mistakes

  • Too few parts close to the limits. Then there are large gaps between the last “−” and the first “+”, and d is overestimated.
  • Reference values not accurate enough. The reference measurement must be much more accurate than the gage, otherwise the sorting blurs.
  • Appraisers know the reference values. Then they rate according to them and not according to the gage.
  • Temperature. With tight tolerances even hand warmth shifts the size. Let parts and gage adjust first.
  • Overlooking gage wear. A worn go side shifts both uncertainty zones. Calibrate the gage itself regularly.

In STATlum

Create an “MSA 7” study on the characteristic and choose “Analytic” or “Consensus”. In the analytic variant you enter the reference value and the ratings 1 (accepted) and 0 (rejected) per part. The evaluation shows all indices of this example.

STATlum: ratings of an MSA 7 study
Ratings of an MSA 7 study: 1 accepted, 0 rejected, one column per trial.
STATlum: results of an MSA 7 study
Results for a limit gage: %GRR, uncertainty zone d and acceptance curve.

Frequently asked questions

When signal detection, when kappa?

Signal detection needs continuous reference values, for example for gages checking a measurable size. Without reference values, as in visual inspection, the agreement is assessed with kappa (MSA 6).

How many parts does signal detection need?

The AIAG example uses 50 parts whose reference values cover both limits. The more parts lie close to the limits, the more precise d becomes.

What are typical acceptance criteria?

Commonly %GRR = d / T below 10 % is acceptable and up to 30 % depending on the application. Requirements of your customer or plant take precedence.

Sources

This article explains the method. It does not replace the guideline that applies to your part or the professional assessment of an actual study.